A biomechanical evaluation of a novel surgical reconstruction technique of the ulnar collateral ligament of the elbow joint

MISSING IMAGE

Material Information

Title:
A biomechanical evaluation of a novel surgical reconstruction technique of the ulnar collateral ligament of the elbow joint
Physical Description:
Book
Language:
English
Creator:
Williams, Nicole
Publisher:
University of South Florida
Place of Publication:
Tampa, Fla
Publication Date:

Subjects

Subjects / Keywords:
Sawbones®
Failure testing
Gap formation
Valgus stability
Ultimate strength
Dissertations, Academic -- Biomedical Engineering -- Masters -- USF   ( lcsh )
Genre:
non-fiction   ( marcgt )

Notes

Abstract:
ABSTRACT: The objective of this thesis is to biomechanically evaluate a novel Double bundle technique for UCL reconstruction designed to accelerate recovery time and minimize gap formation. Excluding UCL surgery, ligament reconstruction procedures typically require an average of 6 months of recovery time. UCL reconstructive surgery requires approximately 1-2 years of recovery time. Valgus instability of the elbow is characterized by attenuation, or frank rupture of the UCL from repetitive and excessive valgus loads. This research compared the valgus stability, gap formation, and ultimate strength that resulted from the cyclic valgus loading at 30 ° of flexion of 3 techniques for reconstruction of the UCL: the Jobe, Docking, and a novel Double bundle procedure. A servocontrolled materials testing machine applied a cyclic valgus load to white cortical Sawbones elbow complex models while a 3D electromagnetic motion tracking system recorded the valgus displacement of the UCL reconstructions. The valgus stability, gap formation, and ultimate strength were measured at 50, 100, 200 and 600 cycles or the cycle at which failure occurred. The mean peak load to failure was 30N for the Jobe reconstructions, and 50N for both the Docking and Double bundle reconstructions. Both the Docking and the Double bundle reconstructions sustained a higher load to failure than the Jobe reconstruction. None of the separate univariate ANOVAs of the biomechanical parameters of each reconstruction were statistically significant. Although there was no statistically significant difference, a small standard deviation in all measured values indicated consistency in testing methodology. The power or sample size is not high enough to state with confidence that statistically there is no difference.
Thesis:
Thesis (M.S.B.E.)--University of South Florida, 2008.
Bibliography:
Includes bibliographical references.
System Details:
Mode of access: World Wide Web.
System Details:
System requirements: World Wide Web browser and PDF reader.
Statement of Responsibility:
by Nicole Williams.
General Note:
Title from PDF of title page.
General Note:
Document formatted into pages; contains 94 pages.

Record Information

Source Institution:
University of South Florida Library
Holding Location:
University of South Florida
Rights Management:
All applicable rights reserved by the source institution and holding location.
Resource Identifier:
aleph - 002021446
oclc - 428440989
usfldc doi - E14-SFE0002589
usfldc handle - e14.2589
System ID:
SFS0026906:00001


This item is only available as the following downloads:


Full Text
xml version 1.0 encoding UTF-8 standalone no
record xmlns http:www.loc.govMARC21slim xmlns:xsi http:www.w3.org2001XMLSchema-instance xsi:schemaLocation http:www.loc.govstandardsmarcxmlschemaMARC21slim.xsd
leader nam 2200385Ka 4500
controlfield tag 001 002021446
005 20101007122208.0
007 cr mnu|||uuuuu
008 090729s2008 flu s 000 0 eng d
datafield ind1 8 ind2 024
subfield code a E14-SFE0002589
035
(OCoLC)428440989
040
FHM
c FHM
049
FHMM
090
R856 (Online)
1 100
Williams, Nicole.
2 245
A biomechanical evaluation of a novel surgical reconstruction technique of the ulnar collateral ligament of the elbow joint
h [electronic resource] /
by Nicole Williams.
260
[Tampa, Fla] :
b University of South Florida,
2008.
500
Title from PDF of title page.
Document formatted into pages; contains 94 pages.
502
Thesis (M.S.B.E.)--University of South Florida, 2008.
504
Includes bibliographical references.
516
Text (Electronic thesis) in PDF format.
3 520
ABSTRACT: The objective of this thesis is to biomechanically evaluate a novel Double bundle technique for UCL reconstruction designed to accelerate recovery time and minimize gap formation. Excluding UCL surgery, ligament reconstruction procedures typically require an average of 6 months of recovery time. UCL reconstructive surgery requires approximately 1-2 years of recovery time. Valgus instability of the elbow is characterized by attenuation, or frank rupture of the UCL from repetitive and excessive valgus loads. This research compared the valgus stability, gap formation, and ultimate strength that resulted from the cyclic valgus loading at 30 of flexion of 3 techniques for reconstruction of the UCL: the Jobe, Docking, and a novel Double bundle procedure. A servocontrolled materials testing machine applied a cyclic valgus load to white cortical Sawbones elbow complex models while a 3D electromagnetic motion tracking system recorded the valgus displacement of the UCL reconstructions. The valgus stability, gap formation, and ultimate strength were measured at 50, 100, 200 and 600 cycles or the cycle at which failure occurred. The mean peak load to failure was 30N for the Jobe reconstructions, and 50N for both the Docking and Double bundle reconstructions. Both the Docking and the Double bundle reconstructions sustained a higher load to failure than the Jobe reconstruction. None of the separate univariate ANOVAs of the biomechanical parameters of each reconstruction were statistically significant. Although there was no statistically significant difference, a small standard deviation in all measured values indicated consistency in testing methodology. The power or sample size is not high enough to state with confidence that statistically there is no difference.
538
Mode of access: World Wide Web.
System requirements: World Wide Web browser and PDF reader.
590
Advisor: William E. Lee III, Ph.D.
653
Sawbones
Failure testing
Gap formation
Valgus stability
Ultimate strength
0 690
Dissertations, Academic
z USF
x Biomedical Engineering
Masters.
773
t USF Electronic Theses and Dissertations.
4 856
u http://digital.lib.usf.edu/?e14.2589



PAGE 1

A Biomechanical Evaluation of a Novel Surgi cal Reconstruction Tec hnique of the Ulnar Collateral Ligament of the Elbow Joint by Nicole Williams A thesis submitted in partial fulfillment of the requirements for the degree of Master of Science in Biomedical Engineering Department of Chemical a nd Biomedical Engineering College of Engineering University of South Florida Major Professor: William E. Lee III, Ph.D. Charles Nofsinger, M.D. Zong-Ping Luo, Ph.D. Date of Approval: July 10, 2008 Keywords: Sawbones , failure testing, gap formation, va lgus stability, ultimate strength Copyright 2008 Nicole Williams

PAGE 2

ACKNOWLEDGEMENTS I would like to thank my committee members for their guidance and support throughout this process. Each of them has e nhanced my critical th inking and scientific reasoning skills and for that I will always be grateful. I extend sincere gratitude to the staff of the Florida Orthopaedic Institute, a nd especially to Charlie Comiskey. Without your dedication and hard work I would not have been able to complete this research. My parents, who have continued to give me spiritual, emotional, and financial support— thank you and I love you very much. To Br ian and Sheena—thanks for staying up late with me and being there to the wire! My extended family and friends—thank you for listening, keeping me updated on the NBA playo ffs, proof reading, editing, and prayer. In closing, I would like to thank the NSF Fl orida Georgia Louis Stokes Alliance for Minority Participation (FGLSAMP) (N SF HRD# 0127675) and Arthrex for their financial support.

PAGE 3

i TABLE OF CONTENTS LIST OF TABLES ............................................................................................................. ii i LIST OF FIGURES ............................................................................................................ v ABSTRACT ..................................................................................................................... viii CHAPTER 1: INTRODUCTION ....................................................................................... 1 1.1 Overview .................................................................................................................. 1 1.2 Clinical Significance ................................................................................................. 1 1.3 Objectives and Hypothesis ........................................................................................ 3 1.4 Limitations ............................................................................................................... 4 CHAPTER 2: BACKGROUND ......................................................................................... 5 2.1 Anatomy and Function of the Human Elbow ........................................................... 5 2.1.1 Bony Anatomy of the Elbow Joint ..................................................................... 5 2.1.2 Articulations of the Elbow Joint ........................................................................ 8 2.1.3 Ligaments of the Elbow Joint ............................................................................ 9 2.1.4 Biomechanics of the Elbow ............................................................................. 13 2.2 The Function of the Ulnar Collatera Lligament ...................................................... 16 2.2.1 Incidences of Injury to the Ulnar Collateral Ligament ......................................................................................................... 17 2.3 Review of Tendon Surgic al Reconstructions .......................................................... 18 2.3.1 Overview .......................................................................................................... 18 2.3.2 Jobe Procedure ................................................................................................. 20 2.3.3 Docking Procedure ........................................................................................... 22 2.3.4 Double Bundle Procedure ................................................................................ 24 2.4 Biomechanics of Tendon Reconstructions ............................................................. 26 2.5 Biomechanics of Intact Ulnar Collateral Ligament ................................................ 28 2.5.1 Valgus Stability ................................................................................................ 28 2.5.2 Gap Formation ................................................................................................. 29 2.5.3 Ultimate Strength ............................................................................................. 30 CHAPTER 3: METHODOLOGY .................................................................................... 31 3.1 Overview ................................................................................................................. 31 3.2 Instrumentation ....................................................................................................... 31 3.2.1 Servocontrolled Materials Testing Machine .................................................... 31 3.2.2 3D Motion Tracking System .............................................................................34 3.3 Specimen Preparation ............................................................................................. 36 3.4 Reconstructive Surgeries ........................................................................................ 38 3.5 Testing Methodology .............................................................................................. 38 3.6 Data Consolidation ................................................................................................. 40

PAGE 4

ii 3.7 Processing and Analysis ......................................................................................... 42 3.7.1 Valgus Stability ................................................................................................ 42 3.7.2 Gap Formation ................................................................................................. 43 3.7.3 Ultimate Strength ............................................................................................. 43 CHAPTER 4: RESULTS .................................................................................................. 44 4.1 Overview ................................................................................................................. 44 4.2 Tendon Simulations ................................................................................................ 44 4.3 Jobe Reconstructed Ulnar Collateral Ligament ...................................................... 48 4.3.1 Valgus Stability ................................................................................................ 48 4.3.2 Gap Formation ................................................................................................. 52 4.3.3 Ultimate Strength ............................................................................................. 55 4.4 Docking Reconstructed Ulnar Collateral Ligament ................................................ 58 4.4.1 Valgus Stability ................................................................................................ 58 4.4.2 Gap Formation ................................................................................................. 62 4.4.3 Ultimate Strength ............................................................................................. 65 4.5 Double Bundle Reconstructed Ul nar Collateral Ligament ..................................... 68 4.5.1 Valgus Stability ................................................................................................ 68 4.5.2 Gap Formation ................................................................................................. 72 4.5.3 Ultimate Strength ............................................................................................. 75 CHAPTER 5: DISCUSSION ............................................................................................ 79 5.1 Overview ................................................................................................................. 79 5.2 Tendon Simulations ................................................................................................ 79 5.3 Comparison of Surgical Reconstructions ............................................................... 81 5.3.1 Valgus Stability ................................................................................................ 82 5.3.2 Gap Formation ................................................................................................. 83 5.3.3 Ultimate Strength ............................................................................................. 84 CHAPTER 6: SUMMARY............................................................................................... 86 CHAPTER 7: RECOMMENDATIONS FOR FUTURE WORK .................................... 88 REFERENCES ................................................................................................................. 90 APPENDICES .................................................................................................................. 9 3 Appendix A: SAS Code ................................................................................................ 94

PAGE 5

iii LIST OF TABLES Table 1: Knot maximum elongation and maximum load and stiffness values for Arthrex Fiberwire #2 and #5 .....................................................45 Table 2: Recorded maximum load a nd manufacturer’s maximum load for Arthrex Fiberwire #2 and #5 ......................................................................45 Table 3: Individual valgus stabi lity and flexibility for the Jobe reconstructed UCL at cycles 50, 100, 200 and 600. ..................................48 Table 4: Mean and standard deviati on values of valgus stability and flexibility for the Jobe rec onstructed UCL at cycles 50, 100, 200 and 600 ................................................................................................49 Table 5: Individual gap formation fo r the Jobe reconstructed UCL at cycles 50, 100, 200 and 600 .......................................................................52 Table 6: Mean and standard deviati on values of gap formation for the Jobe reconstructed UCL at cycles 50, 100, 200 and 600 ..........................53 Table 7: Individual ultimate strength for the Jobe reconstructed UCL at cycles 50, 100, 200 and 600 .......................................................................55 Table 8: Mean and standard deviation values of ultimate strength for the Jobe reconstructed UCL at cycles 50, 100, 200 and 600 ...........................56 Table 9: Individual valgus stabi lity and flexibility for the Docking reconstructed UCL at cycles 50, 100, 200 and 600 ..................................59 Table 10: Mean and standard deviati on values of valgus stability and flexibility for the Docking reconstructed UCL at cycles 50, 100, 200 and 600 ........................................................................................59 Table 11: Individual gap formation for the Docking reconstructed UCL at cycles 50, 100, 200 and 600 ...................................................................62

PAGE 6

iv Table 12: Mean and standard deviati on values of gap formation for the Docking reconstructed UCL at cycles 50, 100, 200 and 600 .....................63 Table 13: Individual ultimate strength for the Docking reconstructed UCL at cycles 50, 100, 200 and 600 ..........................................................65 Table 14: Mean and standard deviation values of ultimate strength for the Docking reconstructed UCL at cycles 50, 100, 200 and 600 .....................66 Table 15: Individual valgus stabilit y and flexibility for the Double bundle reconstructed UCL at cycles 50, 100, 200 and 600 ...................................69 Table 16: Mean and standard deviati on values of valgus stability and flexibility for the Double bundle reconstructed UCL at cycles 50, 100, 200 and 600 .......................................................................69 Table 17: Individual gap formation for the Double bundle reconstructed UCL at cycles 50, 100, 200 and 600 ...................................72 Table 18: Mean and standard deviati on values of gap formation for the Double bundle reconstructed UCL at cycles 50, 100, 200 and 600 ...........73 Table 19: Individual ultimate strength for the Double bundle reconstructed UCL at cycles 50, 100, 200 and 600 ...................................75 Table 20: Mean and standard deviation values of ultimate strength for the Double bundle reconstructed UCL at cycles 50, 100, 200 and 600 ...........76 Table 21: Depicts the mean displacement at each load step for the Jobe, Docking, and Double bundle reconstructions ............................................83

PAGE 7

v LIST OF FIGURES Figure 1: (A) Diagrammatic AP view of elbow joint .................................................6 Figure 2: True lateral ra diograph of the humerus ........................................................7 Figure 3: Osseous congruencies of ulna-radius—anterior and lateral aspects ..........................................................................................................8 Figure 4: Articulating surface (tra ced by black lines) of the (A) humerus and (B) ulna and ra dius of the elbow joint ....................................9 Figure 5: A picture of the ligament complexes of the elbow (A) left elbow joint, showing anterior and ulnar collateral ligaments. ...................10 Figure 6: Cam effect of the differe nt bundles of the ulnar collateral ligament (UCL) varies based on the degree of elbow flexion ...................11 Figure 7: The anterior bundle, posteri or band, and transverse ligament of the medial ligamentous complex of the elbow ..........................................12 Figure 8: Anatomy of the late ral ligaments of the elbow ..........................................13 Figure 9: Medial elbow muscles insert ions (A) flexor carpi radialis (B) flexor digitorum superficia lis (C) pronator teres flexor. .....................15 Figure 10: Valgus stress placed on the ulnar collateral ligament during arm cocking and acceleration phase in baseball pitching .................................17 Figure 11: Magnetic resonance imaging scan showing torn ulnar collateral ligament......................................................................................................18 Figure 12: Procedural depicti on of the Jobe technique. ..............................................21 Figure 13: Procedural depiction of the Docking technique. ........................................23 Figure 14: Procedural depiction of the Double bundle technique ...............................25

PAGE 8

vi Figure 15: Imaging studies. .........................................................................................29 Figure 16: Test Resources 800L series servocontrolled ma terials testing machine ......................................................................................................33 Figure 17: Diagram of testing apparatus used in study ...............................................34 Figure 18: 3D electromagneti c motion tracking system ..............................................35 Figure 19: Straight pull test of Arthrex Fiberwire #5 ..................................................36 Figure 20: Disarticulated Sawbones w ith the humerus and forearms potted in PVC tubes with Bondo ..........................................................................37 Figure 21: Organizational flow chart depicting the testing methodology ...................39 Figure 22: Organizational flow char t depicting data consolidation ............................40 Figure 23: Valgus angle equation ................................................................................42 Figure 24: Moment calculation ...................................................................................43 Figure 25: Graphs depicting the stiffn ess of looped suture specimens (A) Fiberwire #2 (B) Fiberwire #5 ...................................................................46 Figure 26: Graphs depicting the stiffness of suture specimen s with a straight pull test (A) suture specimen 1 (B) suture specimen 2 .............................47 Figure 27: Valgus stability of Jobe UC L reconstructions for (A) individual specimens and (B) and the mean of specimens .........................................50 Figure 28: Flexibility of Jobe UCL reconstructions fo r (A) individual specimens and (B) the mean of specimens ................................................51 Figure 29: Gap formation of Jobe UCL reconstruction for (A) individual specimens and (B) the mean of specimens ................................................54 Figure 30: Ultimate strength of Jobe UC L reconstructions for (A) individual specimens and (B) the mean of specimens ................................................57 Figure 31: Number of cycles to fail ure for each Jobe UCL reconstruction ................58 Figure 32: Valgus stability of Docking UCL reconstructions for (A) individual specimens and (B) the mean of specimens ........................60

PAGE 9

vii Figure 33: Flexibility of Docking UCL reconstructions for (A) individual specimens and (B) the mean of specimens ................................................61 Figure 34: Gap formation of Docking UC L reconstructions for (A) individual specimens and (B) the mean of specimens ................................................64 Figure 35: Ultimate strength of Docking UCL reconstructions for (A) individual specimens and (B) the mean of specimens ........................67 Figure 36: Number of cycles to fail ure for each Docking reconstruction ...................68 Figure 37: Valgus stability of D ouble bundle UCL reconstructions for (A) individual specimens and (B) the mean of specimens ........................70 Figure 38: Flexibility of Double bundle UCL reconstructions for (A) individual specimens and (B) the mean of specimens ........................71 Figure 39: Gap formation of Doubl e bundle UCL reconstructions for (A) individual specimens and (B) the mean of specimens ........................74 Figure 40: Ultimate strength of Doubl e bundle UCL reconstructions for (A) individual specimens and (B) the mean of specimens ........................77 Figure 41: Number of cycles to failure for each Double bundle reconstruction .............................................................................................78 Figure 42: Valgus stability comparis on of the Jobe, Docking and Double bundle reconstructions at cycles 50, 100, 200 and 600 ..............................82 Figure 43: Gap formation comparison of the Jobe, Docking and Double bundle reconstructions at cycles 50, 100, 200 and 600 ..............................84 Figure 44: Ultimate strength comparison of the Jobe, Docking and Double bundle reconstructions at cycles 50, 100, 200 and 600 ..............................85 Figure 45: Mean number of cycles to failure of the Jobe, Docking, and Double bundle reconstructions ..................................................................85

PAGE 10

viii A BIOMECHANICAL EVALUATION OF A NOVEL SURGICAL RECONSTRUCTION TECHNIQUE OF THE ULNAR COLLATERAL LIGAMENT OF THE ELBOW JOINT Nicole Williams ABSTRACT The objective of this thesis is to biomechanically ev aluate a novel Double bundle technique for UCL reconstruction designed to accelerate recovery time and minimize gap formation. Excluding UCL surgery, ligament rec onstruction procedures typically require an average of 6 months of recovery tim e. UCL reconstructive surgery requires approximately 1-2 years of recovery time. Val gus instability of the elbow is characterized by attenuation, or frank rupture of the UCL fr om repetitive and excessive valgus loads. This research compared the valgus stabil ity, gap formation, and ultimate strength that resulted from the cyclic valgus loading at 30 of flexion of 3 techniques for reconstruction of the UCL: the Jobe, Do cking, and a novel Double bundle procedure. A servocontrolled materials testing machin e applied a cyclic valgus load to white cortical Sawbones elbow complex models while a 3D electroma gnetic motion tracking system recorded the valgus di splacement of the UCL reconstr uctions. The valgus stability, gap formation, and ultimate strength were m easured at 50, 100, 200 and 600 cycles or the cycle at which failure occurred. The mean p eak load to failure was 30N for the Jobe reconstructions, and 50N for both the Dock ing and Double bundle re constructions. Both the Docking and the Double bundle reconstructions sustained a higher lo ad to failure than the Jobe reconstruction. None of the sepa rate univariate ANOVAs of the biomechanical

PAGE 11

ix parameters of each reconstruction were statis tically significant. Although there was no statistically significant difference, a small st andard deviation in all measured values indicated consistency in testing methodology. The power or sample size is not high enough to state with confidence that st atistically there is no difference.

PAGE 12

1 CHAPTER 1: INTRODUCTION 1.1 Overview The past few decades have shown a substa ntial increase in the interest in ulnar collateral ligament (UCL) rec onstructive surgery. By 2001, 75 major league pitchers had undergone UCL reconstructive su rgery which is approximately one out of every nine pitchers. A recent retrospective study over the period of 1988-2003 revealed in addition to an overall dramatic increase in rec onstructive surgery, a 50% increase in UCL reconstruction in high school athlet es (aged 15-19). (Langer, 2006) The following subsections will briefly but fully detail the clinical significance, objectives and hypotheses, and the lim itations of this research study. 1.2 Clinical Significance UCL injuries are common and are most frequently seen in overhead throwing athletes such as baseball pitchers and tenni s players. Early recognition and treatment are required to ensure the possibility of return to pre-injury leve ls of participation. UCL reconstruction described by Jobe et al is a widely accepted procedure in the treatment of symptomatic elbow instability. Th is process involves a tendon graft, usually the palmaris longus, pulled through bone tunnels in the medial epicondyle of the humerus and ulnar coronoid process in a figure-eight fashion and th e transposition of the ulna

PAGE 13

2 nerve. The main complication of this method is related to the ul nar nerve. (Hechtman, 1998, Tjin, 1998) Excluding UCL reconstruction surgery, ligament reconstruction procedures typically require an average of 6 months of recovery time. Upon a ppropriate recognition of the need for reconstruction, the optimal length of recovery time by the Jobe or Docking reconstructions is approximately 12 months. (Nassab, 2006) Conway and colleagues reported that 68% of their patients returned to pre-injury status with a mean recovery time of 12 months. Current biomech anical studies show that all ligament reconstructions are inferior to the native medial ligame nt complex and show no clear biomechanical advantage of one recons truction over another. (Nassab, 2006) Additionally, clinical studies have shown that cutaneous ne rve injury and ulnar nerve neuropathologies are by far the most comm on complications associated with this procedure and may occur in up to 25% of cases. (Leibman M. 2002) One of the major concerns with the muscle-splitting approach is the proximity of the posterior ulnar drill hole to the ul nar nerve in the figure-eight and Docking reconstructions. There is also the concern th at when sutures and tendons are being passed through these holes, fixation could be lost if the ulnar bridge is compromised. (Armstrong, 2005) Biomechanical and kinematic studies have suggested that a more anatomical reconstruction of the UCL may provi de improved long term outcomes. I hypothesize that the use of a doubl e bundle in a UCL reconstruction will anatomically mimic the reciprocal tensioning and flexing of the an terior and posterior bands of the UCL. Double bundle procedures have been used in the reconstruction of other multi-bundle ligaments an example of which is the anterior cruciate ligament (ACL) Yagi et al. reported that anatomic double bundle ACL r econstruction restores knee kinematics more closely to normal than does a single bundle ACL reconstruction. In particular, anatomic double bundle ACL reconstr uctions result in be tter rotational knee stability than does a single bundle ACL rec onstruction. (Chhabra, 2006) This research study will compare the biomechanical pr ofiles of the Jobe and Docking UCL reconstruction techniques with a novel double bundle procedure. These results will offer orthopaedic clinicians insight into the effects of a more anatomic reconstruction.

PAGE 14

3 1.3 Objectives and Hypothesis The objective of this thesis is to make a quantitative assessment of the biomechanical profiles of two industry st andard (Jobe and Docking) and one novel double bundle UCL recons truction technique. I hypothesize that the use of a double bundle in a UCL reconstruction of the intermediate fibers will maximize the isometry within the anterior oblique ligament and allow the accurate reproduction of the tens ioning of each bundle of the UCL to restore native elbow stability. This assessment is based on the following defined goals: 1) The valgus stability of the elbow as a function of cyclic valgus loading 2) The elongations of the UCL reconstruction wh en subjected to a valgus displacement at a 30 angle of flexion 3) The ultimate strength of the UCL r econstructions at the instance of failure 4) The stiffness of th e UCL reconstructions The design of this study will provide insight into an appropriate surgical method that mimics the anatomy and functional ity of medial el bow stability.

PAGE 15

4 1.4 Limitations Sawbone elbow models will be used in th e experiment to remove the variability associated with cadaver studies. These models are inexpensive, readily available, anatomically accurate, and physically iden tical to their cadaveric counterpart. Many studies have used Sawbones for evaluating mechanical properties of various constructs or for holding abilities of internal fixation. Langdsman and Chang studied the validity of Sawbone models “when mechanical integrity of the model was critical for evaluating the outcome.” They found that, although Sawbones do not simulate the mechanical properties seen in cadaveric bone, they still can be used in studies in which the relative stability is being assessed. They believed that the relative values were significant, even if the absolute values were not. In this study, a uniform platform for comparison of UCL reconstructions was crea ted by using Sawbones, thus resulting in meaningful clinical data. (Khuri, 2003) Another shortcoming of this study is that it represents results at time=0, when no scarring or healing has occurred, and is impo ssible to reach clinical loads. Specimens were only loaded in supination at 30 of flex ion, whereas instability can present itself as a spectrum throughout the arc of elbow flexion and extension. (Armstrong, 2005)

PAGE 16

5 CHAPTER 2: BACKGROUND 2.1 Anatomy and Function of the Human Elbow The elbow is a synovial hinge joint that is composed of various skeletal and soft tissue constraints that contribu te to its anatomy and functi on. Skeletal anatomy such as the olecranon/olecrenon fossa ar ticulation provide primary stability of less than 20 or greater than 120 of elbow flexion. Soft tissue constraints provide the elbow with dynamic and static stability during the overhead arc of motion (20-120 ). There are three primary functions of the elbow: 1) To serves as a component joint of the lever arm in positioning the hand 2) A fulcrum for the forearm lever 3) A weight bearing join t in patients using crut ches (Miller, 1992) 2.1.1 Bony Anatomy of the Elbow Joint The bony anatomy of the elbow consists of three bones: the humerus which is located in the upper part of the arm, and the ulna and radius located in the forearm. (Marieb, 2004) Each bone posse sses characteristic osseous c ongruencies that contribute to the functioning of the elbow. The distal end of the humerus is char acterized by two condyles the medial trochlea and the lateral capitellum. (Marieb, 2004) These condyles ar ticulate with the radius and the ulna respectiv ely. The elbow rotates virtually around a single axis that coincides with the condylotrochl ear axis. (Dumontier, 2006)

PAGE 17

6 Figure 1: (A) Diagrammatic AP view of elbow joint. (B) Diagrammatic lateral view of elbow joint. Note that the elbow is slightly twisted in respect of the axis of the ulna. (Dumontier, 2006) These condyles are bordered by the medial and lateral epicondyles which are the insertion sites of the common flexor and ex tensor tendons from whic h several flexor and extensor muscles originate. The supracondyl ar ridge lies directly above these condyles. The ulnar nerve runs behind the medial ep icondyle. Superior to the trochlea on the anterior and posterior surfaces is the co ronoid fossa and olecranon fossa respectively. These depressions allow the corresponding ol ecranon and coronoid processes of the ulna to move freely while the elbow is engage d in flexion and extension. On a lateral radiograph of the humerus, the centers of th e three circles formed by the edge of the condyles, the ulnar groove, and the medial lip of the trochlear coincide with the elbow’s axis of flexion and extension. (Dumontier, 2006)

PAGE 18

7 Figure 2: True lateral radiograph of the humer us. The centers of the three circles formed by the edge of the condyle, the ulnar groove, a nd the medial lip of the trochlea coincide; this point is the flexion-extension ax is of the elbow. (Dumontier, 2006) The ulna is primarily responsible for fo rming the elbow joint with the humerus. Its proximal end is reminiscent of the adju stable end of a monkey wrench, where the olecranon and the coronoid processes are separated by the trochlear notch. These processes hold the trochlear of the humerus and allow the forearm to flex and extend. When the forearm is in full extension th e olecranon process is “locked” into the olecranon fossa causing the elbow to hyperexten d. On the lateral side of the coronoid process the radial notch (a small depression) articulates with the head of the radius. (Marieb, 2004) The head of the radius is similar to the head of a nail; the superior surface of which is concave and articulates with the capitellum of the humerus Medially, the head articulates with the radial notch of the ulna. Th e radial turbosity is located inferior to the head and is the insertion site of the biceps mu scle. Distally, the radius has a medial ulnar notch that articulates with the ulna, a nd a lateral styloid process. (Marieb, 2004)

PAGE 19

8 Figure 3: Osseous congruencies of ulna-radius —anterior and lateral aspects. (Thinnes, 2006) 2.1.2 Articulations of the Elbow Joint The elbow joint is composed of thre e articulations: the humeroulnar, the humeroradial, and of lesser importance for stab ility and motion the proximal radio-ulnar. Composed of the trochlear of the humerus and the trochlear notch of the ulna the humeroulnar joint is the largest and most st able of the articulati ons of the elbow. Its stability is dependent on the stability of the medial collateral ligament. The humeroradial joint lies lateral to the humeroulnar joint. It is composed of the radial head and the capitellu m of the humerus. This osseous congruency prevents proximal migration of the radius throughout th e full arc of flexion and extension of the elbow. The proximal radio-ulnar joint is compos ed of the radial notch and the radial head. (Miller, 1992)

PAGE 20

9 Figure 4: Articulating surface (traced by black li nes) of the (A) humerus and (B) ulna and radius of the elbow jo int. (Thinnes, 2002) 2.1.3 Ligaments of the Elbow Joint The static soft tissue stabilizers of the elbow joint consist of the medial and lateral ligamentous complexes and the anterior and posterior ligam ents. (Safran, 2005) Additionally, an articular capsule extends inferi orly from the humerus to the ulna and to the annular ligament surrounding the head of the radius. The thinness of the articular

PAGE 21

10 capsule on the anterior and pos terior surfaces of the elbow allow ease of flexion and extension with little resist ance. (Marieb, 2004) The medial and lateral complexes resist lateral (valgus and varus forces) motion. Figure 5: A picture of the ligament complexes of the elbow (A) left elbow joint, showing anterior and ulnar collateral ligaments. (B) Left elbow joint, showing posterior and radial collateral ligaments. (Gray’s Anat omy of the Human Body, WEB) The medial ligamentous complex or th e UCL is subdivided into the anterior oblique ligament (AOL), the posterior obliq ue ligament (POL), and the transverse ligament (TL). The AOL and POL originate from the central portion of the anteroinferior portion of the medial epicondyle. Due to the location of its origin, a cam effect is produced in which ligament tension in creases with increasing flexion. A B

PAGE 22

11 Figure 6: Cam effect of the different bundles of the ulnar collateral ligament (UCL) varies based on the degree of elbow flexion. (Safran, 2005) Composed of thick parallel fibers of approximately 4 to 5 mm, the AOL is the strongest of the complex and is generally thought of as the primary restraint and stabilizer. The AOL inserts along the medial aspect of the coronoid process and is functionally composed of an anterior, central and poste rior band. The anterior band (AB) is taut during the first 60 of flexion and the posteri or band (PB) is taut between 60 and 120 of flexion. This provides a recipro cal function in resisting val gus stress during flexion and extension. The central band is isometric th roughout flexion and extension. (Safran, 2005,

PAGE 23

12 Nassab, 2006) The POL inserts in the medial ma rgin of the semilunar notch and is fanshaped with a width of 5 to 6 mm in the middle of the fan-shaped segment. The TL or Cooper’s ligament consists of horizontal ca psular fibers between the coronoid and the olecranon and does not cont ribute significantly to me dial elbow stability. Figure 7: The anterior bundle, posterior ba nd, and transverse liga ment of the medial ligamentous complex of the elbow. (Thinnes, 2006) The lateral collateral ligament (LCL) is composed of the lateral radial collateral ligament (RCL), annular ligament (AL), late ral ulnar collateral ligament (LUCL), and accessory lateral collateral ligament (ALC L). The AL originates and inserts on the anterior and posterior margins of the lesser sigmoid notch of the ulna. The RCL inserts into the AL and the LUCL inserts via a br oad insertion into the proximal ulna. When present, the ALCL originates from the supinato r crest of the ulna a nd is thought to assist the AL in resisting varus stress. The LUCL originates from th e lateral epicondyle and inserts on the supinator crest of the ulna.

PAGE 24

13 Figure 8: Anatomy of the lateral lig aments of the elbow. (Safran, 2005) The anterior and posterior ligaments play a minor role in the stabilization of the elbow. The anterior ligament is composed of lateral oblique fibers. The posterior ligament is composed of tran sverse and oblique fibers. 2.1.4 Biomechanics of the Elbow The range of motion (ROM) of the el bow includes a 0-150 arc of flexionextension that functionally opera tes from 30-130 with an axis of rotation centered at the trochlea. The ROM also includes pronation and supination at 80 and 85 respectively, both of which function at 50 . The axis of pronation and supination is a line from the capitellum through the radial head and to th e distal ulna. The normal carrying angle for males and females is 7 and 13 respectively. The short lever arms of the forces acting around the elbow are inefficient and result in large joint reaction forces that subject the elbow to degenerative changes. Flexion is primarily through the biceps while ex tension is primarily through the triceps. Pronation is accomplished with muscles in the pronators class (teres and quadratus) and

PAGE 25

14 supination is accomplished through the supinato r. Varus-valgus rotational motions result in elbow instability and are restricted by th e medial and lateral ligamentous complexes. The greatest resistance to rotation occurs on th e medial side of the elbow. In the MCL the AOL is taut throughout the arc of flexion-extension while the POL is taut only during flexion. The MCL contributes 54% of the re sistance to valgus stress during 90 of flexion. The remaining resistance is supplied by the shape of the articular surfaces and the anterior capsule. (Buckwalter, 2000) During elbow flexion the maximum is ometric force of the flexors is approximately 40% greater than the isomet ric force of the extensors. The average maximum torque strength for elbow extensi on in men and women is 4 kgm and 2 kgm respectively. Measurements during forearm prona tion and supination indicate that there is a linear relationship between strength and forearm rotation. The average torque of supination exceeds that of pronation by 15-20% in men and women throughout a variety of shoulder-elbow positi ons. (Buckwalter, 2000) Soft tissue constraints an d bony articulations provide stability in different ratios depending upon the position of the elbow dur ing flexion and extension. (Nassab, 2006) The medial soft tissue constraints of the elbow are subjected to tensile forces that result in valgus stress. In 1983 the research of Morrey et al. showed that the medial soft tissues that resist valgus forces c ontribute different amounts depe nding on the angle of flexion. In full extension, approximately one third of valgus force was resisted by the UCL (31%), one third by the anterior capsule (38%), a nd one third by the bony geometry (31%). At 90 of flexion, the UCL substantially increase d its load of the re sistance to 54%, while the capsule’s contributions were reduced to 10% and the resistance of the bony geometry remained unchanged at 36%. (Safran, 2005) The muscular forces of the elbow confer stability by compressing the joint surfaces against each other. The capacity of a muscular structure to contribute to elbow stability is a measure of the joint position and a balance of the muscles that cross the elbow. Muscles that primarily provide flexionextension are unable to provide significant varus-valgus stability. Hyperextension of th e elbow is resisted by the anterior muscles based on both their bulk and dynamic contributions.

PAGE 26

15 The stabilizing contributions of the fo rearm muscles on the medial side of the elbow have been the subject of much researc h. The medial muscles include: the pronator teres, flexor digitorum supe rficialis, and flexor carpi ra dialis. The muscles provide stability by the application of a varus moment to the elbow and thus can resist valgus force regardless of the forearm position. (Saf ran, 2005) Other muscul ar structures are associated with the static, dynamic, and the posterolateral stability of the elbow. These muscles include: fascial bands, and intermuscu lar septae from the extensor digitorum communis and extensor digitorum quinti. Th ese bands and septae tighten in supination. Figure 9: Medial elbow muscles insertions (A) flexor carpi ra dialis (B) flexor digitorum superficialis (C) pronator teres flexor (Musculoskeletal Radiology, WEB) C B A

PAGE 27

16 In addition to the research of medial so ft tissue stabilization, Morrey et al. studied the resistance to varus stress in both the fully extended and 90 flexed elbow. At full extension, nearly half of the stability is conferred by the lateral soft tissues (14% LCL and 32% by the capsule) and 55% is conferred by the bony geometry. At 90 of flexion, 75% of the resistance to varu s stress is contributed by the bony articulations while the remaining resistance is provided by the LCL (9%) and the capsule (13%). (Safran, 2005) Several independent studies have shown that the LUCL is the primary constraint in the LCL resisting rotatory instability. (Safran, 2005) Additionally, another study by Olsen et al. has shown that rotatory instability is al so related to the amount of elbow flexion. The increased degrees of rotatory laxity which cu lminated in the severance of the LCL, is a result of increased degrees of flexion. This effect was noted maximally at 110 of elbow flexion which produced 20.6 of laxity in forced external rota tion. (Safran, 2005) 2.2 The Function of the Ulnar Collateral Ligament The UCL functions as the pr imary restraint against val gus loading of the elbow due to its location. (Nassab, 2006) The eccentric location of the humeral origin of the AOL with respect to the flexion-extension axis provides stability throughout the ROM. The insertion of the AOL is into the coronoid of the ulna, giving it a mechanical advantage in controlling valgus forces. Fo r a detailed explanati on of the individual contributions of the AOL, POL, and TL to the overall function of the UCL refer to section 2.1.3 Ligaments of the elbow.

PAGE 28

17 2.2.1 Incidences of Injury to the Ulnar Collateral Ligament UCL insufficiency is caused by medial el bow pain and dysfunction that is usually seen in a throwing athlete. The mean valgus pitch of an adult is 64 Nm. The ultimate valgus torque of the UCL is 33 Nm causi ng the UCL complex to approach maximum torque with every pitch. (Lange r, 2006) In kinetic studies, tension on the UCL during the acceleration phase of pitching is between 64 and 120 Nm. When measured directly in human cadaveric studies the ultimate load to failure of the UCL was between 22.7 and 33 Nm. The biomechanics of throwing indicate that a significant valgus lo ad is placed on the elbow resulting in the strain of the anterior band of the UCL. These forces can cause acute rupture or create recurring microtears which result in the eventual erosion of the ligament. Valgus instability of the elbow is characterized by an opening greater than 2 to 3 mm between the coronoid and the trochlea. (Prud’homme, 2008) Figure 10: Valgus stress placed on the ulnar co llateral ligament dur ing arm cocking and acceleration phase in baseball p itching. (Elattrache, 2001)

PAGE 29

18 Figure 11: Magnetic resonance imaging scan showing torn ulnar collateral ligament. Arrows represent torn ends of ligament. (Elattrache, 2001) 2.3 Review of Tendon Surgical Reconstructions The following section is an overview of the Jobe and Docking UCL surgical reconstruction techniques. Additionally, this section reviews the relevant literature relating to double bundle surg ical reconstruction. 2.3.1 Overview The current definition of surgical recons truction success is defi ned as the ability of an athlete to return to a pre-injury level of play for at least one year. The success rate of UCL reconstructions va ries between 63% and 97% depending upon the technique employed. (Langer, 2006) However, biomech anical studies of UCL reconstruction techniques show that all lig ament reconstruction techniques are inferior to the native UCL. (Nassab, 2006) UCL reconstruction requi res a tendon graft which is usually the palmaris longus. Typically 15-17 cm is need ed. Biomechanical studies have shown that the palmaris has a higher ultimate failure load (357 N) than that of the AOL (260 N).

PAGE 30

19 Despite minimal morbidity associated with its removal, the surgeon must avoid damaging the adjacent median nerve. (Langer, 2006) Additional tendon graft choices include: the gracilis, semitendinosous, and the patellar tendon. (Prud’homme, 2008) The surgical reconstruction of the UCL as described by Conway et al. and Jobe et al. is an industry standard procedure in th e treatment of medial elbow instability. The graft is pulled through bone tunnels in the medial epicondyle of the humerus and the ulnar coronoid process in a figure-eight fa shion followed by transposition of the ulnar nerve. In 1986, Dr. Jobe published the results of a 2 year follow up study. 63% of 16 elite throwing athletes were able to return to a pr e-injury level of play for at least 1 year. Despite the success of this procedure there wa s an alarmingly high rate of complications (31%). The primary complication of this proced ure is related to the transposition of the ulnar nerve. In a subsequent follow up st udy in 1992 with a reconstruction patient population of 56, 68% of patients returned to a pr e-injury level of play after a mean of 12 months of recovery. Postoperative ulnar ne rve dysfunction was pres ent in 21% of the patients. (Langer, 2006) In 1996 Altchek et al. used a muscle splitting approach to modify the Jobe procedure. Modification of th is procedure known as the Do cking technique reroutes the graft anteriorly to avoid tran sposition of the ulnar nerve. (Hechtman, 1998, Langer, 2006) Unlike the figure-eight position of the graft in the Jobe technique, the graft in the Docking technique is placed in a triangular configuration through a humeral tunnel and bone punctures and subsequently tied over a bone bridge. In a retrospective study conducted by Rohrbough et al. 33/36 (92%) patients returned to a pre-injury level of play for at least 1 year. I hypothesize that the use of a doubl e bundle in a UCL reconstruction of the intermediate fibers will maximize the isom etry within the AOL and allow the accurate reproduction of the tensioning of each bundle of the UCL to restore native elbow stability. (Ahmad, 2003, Lee, 2005)

PAGE 31

20 2.3.2 Jobe Procedure The procedure described in the followi ng section has been adapted from the independent studies performed by Paletta et al. and Conway et al. The sites of attachment of the AOL are located on the humerus and th e ulna. Bone tunnels were created in the proximal ulna and medial epicondyle of the hum erus using a motorized drill and a drill guide. The tunnels are placed so that the graft material will not rub against the epicondyle. The ulnar and humeral bone tunnels were created with a 3-mm and 4-mm drill respectively. A single hole is drilled superiorly into the ul na so that the bony bridge correlates with the attachment site of the AOL. The humeral tunnel is Y-shaped with a shor t straight limb at the insertion site of the UCL and 2 branched limbs of identical diam eter. The first hole is drilled anteriorly at the site of the origin the UCL and the second posteriorly at approximately a 30 angle from the first. The third hole is drilled in th e posterior aspect of th e epicondyle so that it is collinear with the first tunnel. (Paletta 2006, Conway, 1992) Arthrex Fiberwire #5 is passed through the bone tunnels in a figureeight fashion and tens ioned in a Krackow locking pattern.

PAGE 32

21 Figure 12: Procedural depiction of the Jobe technique. (Thinnes, 2006)

PAGE 33

22 2.3.3 Docking Procedure The procedure in the following section ha s been adapted from the research of Altchek et al. The sites of attachment of th e AOL are located on the humerus and the ulna. Bone tunnels were created in the proximal ulna and medial epicondyle of the humerus using a motorized drill and a drill guide. Ulna r tunnels were made anterior and posterior to the sublime tubercle by using a 3-mm drill to create a 2-cm bridge between the tunnels. The tunnels were connected using a small, cu rved curette. The humeral tunnel position was located in the anterior half of the medial epicondyle. A longitudinal tunnel was created up the axis of the medial epicondyle to a depth of 15 mm by using a 4-mm drill. With the use of a dental drill with a sma ll bit, two small exit punctures separated by 5 mm to 1 cm were created to allow suture passage from the primary humeral tunnel. Arthrex Fiberwire #5 was then passed through the ulna from ante rior to posterior. One end of the Fiberwire was passed into the humeral tunnel a nd one of the small superior humeral punctures. With this first limb securely docked in the humerus, the second half of the Fiberwire is visually m easured to estimate the length needed for tensioning the other limb. This end of the graf t is docked securely in the humeral tunnel exiting the small puncture holes. Once the surgeon is satisfied with the Fiberwire tensioning, both sets are tied over the bony bridge on the humeral epicondyle. (Rohrbough, 2002)

PAGE 34

23 Figure 13: Procedural depiction of th e Docking technique. (Thinnes, 2006)

PAGE 35

24 2.3.4 Double Bundle Procedure The Double bundle procedure used is an EndoButton procedure on both the humerus and the ulna with the two tunnels on the ulna converging to a single exit laterally. This will allow vary ing the placement of the tunnels and the tensioning of each ligament, anterior and posterior. The current technique uses only the standard positions to compare the biomechanics of the EndoButt on fixation to the sta ndard reconstructions. The sites of attachment of the AOL are located on the humerus and the ulna. Bone tunnels were created in the proximal ulna a nd medial epicondyle of the humerus using a motorized drill and a drill guide. Ulnar tunnels were made anterior and posterior to the sublime tubercle by using a 3-mm drill to create a 2-cm bridge. The tunnels were connected using a small, curved curette. Th e humeral tunnel position was located in the anterior half of the medial epicondyle. A l ongitudinal tunnel was cr eated up the axis of the medial epicondyle to a depth of 15 mm by using a 4-mm drill. With the use of a dental drill with a small bit, two small exit punctures separated by 5 mm to 1 cm were created to allow suture passage from the primary humeral tunnel. Arthrex Fiberwire #5 was then passed thr ough the ulna from ante rior to posterior. One end of the Fiberwire was passed into the humeral tunnel a nd one of the small superior humeral punctures. With this first limb securely docked in the humerus, the second half of the Fiberwire is visually m easured to estimate the length needed for tensioning the other limb. This end of the graf t is docked securely in the humeral tunnel exiting the small puncture holes. Once the surgeon is satisfied with the Fiberwire tensioning, both sets are tied over the bony br idge on the humeral epicondyle with an EndoButton.

PAGE 36

25 Figure 14: Procedural depiction of the Double bundle technique.

PAGE 37

26 2.4 Biomechanics of Tendon Reconstructions The success rates of UCL reconstructions vary from 63% to 97% depending on the technique employed. Complication rates are reportedly less than 10%. Despite the variability in the techniques, the unifying elements are th e decreased dissection of the flexor-pronator mass and the decreased handling of the ulnar nerve, that have lead to advantageous outcomes. (Langer, 2006) Bi omechanical studies have shown that UCL reconstructions are inferior to the native UCL complex. Ther e have been 5 major studies in the past decade that have focused on the biomechanical properties of UCL reconstructions. Hechtman et al. compared the biomechanical properties of the native UCL complex with both the Jobe technique and a le ss invasive reconstr uction procedure that replaced the traditional bone tunnels with bone anchors. Elbow specimens were loaded to failure in a materials testing machine at a fi xed angle of 30 . The results of this study indicated that the bone anchors were better able to mimic the strain patterns of the AOL and POL during flexion and extension. Howe ver, the intact ligament was still significantly stronger than either reconstruction. A novel interference screw technique de signed to minimize soft tissue dissection, injury to the ulnar nerve, and ease of fixa tion was researched by Ahmad et al. Intact, disrupted and reconstructed ligaments were te sted both kinematically and to failure. This technique though significantly inferi or in overall stiffness to the native UCL, was able to nearly replicate the ultimate moment and valgus stability of the native complex. This study did not incorporate cyclic loading whic h would assess the fixation characteristics during the early postoperative course. Armstrong et al. compared the initial st rength of the Jobe, Docking, EndoButton, and interference screw UCL reconstruction t echniques with each other and the native complex. The respective reconstructions were cy clically loaded to failure in a stepwise manner. The initial strength and mean disp lacement of the graft at varying loads was significantly lower than the native UCL complex.

PAGE 38

27 The biomechanical parameters of the Jobe and Docking techniques with an increased graft surface area were analyzed by Paletta et al. Elbows were potted and tested using a servohydraulic materials testing machine to apply a valgus moment at 30 of elbow flexion. Both the native ligament and the Docking construct showed similar values for maximal moment to failure. Howe ver, the native ligament complex exhibited a substantially better biomechanic al profile than either re construction technique. More specifically, the native complex exhibited a hi gher stiffness and lower strain at maximal moment. McAdams et al. evaluated th e effect of cyclic valgus loading on the Docking and bioabsorbable interference screw techniques. A cyclic valgus load was applied to elbows and the valgus angle was measured at vary ing cycles. The bioabsorbable interference screw technique resulted in a smaller val gus angle widening as compared with the Docking technique. Generally, there is no clear biomechanic al superiority of one reconstruction technique over the other. These studies have shown an ability to replicate native maximal moment to failure but fail to approach the sti ffness or clinically rele vant strength of the native complex. The purpose of this study is to evaluate the biomechanical parameters of a novel Double bundle UCL recons truction technique in comp arison with the Jobe and Docking techniques. Elbow constructs will by cy clically loaded to failure to measure: valgus stability, gap formation and ultimate strength.

PAGE 39

28 2.5 Biomechanics of Intact Ulnar Collateral Ligament The UCL’s ability to function as the prim ary valgus support of the elbow is due to its location and structure. Of the three ligaments that comprise the UCL complex, the AOL provides a significant amount of the rest raint to valgus stress in the elbow. The medial elbow is subjected to extreme val gus stresses during throwing. These forces are the greatest during the late cocking and early accelera tion phase of pitching. The biomechanics of the UCL have been the subj ect of several clinical and biomechanical studies. In the following subsections biomechani cal parameters including: valgus stability, gap formation and ultimate strength will be discussed in further detail. 2.5.1 Valgus Stability Morey et al. quantified the c ontributions of different stru ctures to valgus stability as a function of flexion angle. In full extens ion approximately one third of valgus forces are resisted by the UCL, anterior capsule, and bony anatomy respectively. In contrast, during 90 of flexion the UCL increased its valgus stabilizing contributions to 56% while the anterior capsule was reduced to 10% and the bony anatomy contributions remained generally the same. Additionally, wh en the entire UCL was removed combined with radial head deficiency, gross valgus instability and in ternal rotation were observed. Complete removal of the radial head coupled with resection of the UCL lead to elbow sublaxation at 120 of elbow flexion. In this research study the valgus stability of the reconstructe d elbow as a function of cyclic valgus loading was evaluated. A stronger UCL reconstr uction should reduce the valgus angle in response to the ap plication of a valgus moment.

PAGE 40

29 2.5.2 Gap Formation A clinical indication of UCL insufficienc y is a medial joint opening of greater than 2 mm in response to valgus loading. (E lattrache, 2001) Rajike et al. assessed the differences in stress radiographs of an in jured and uninjured patient population. Their results showed that a gapping of less than 0.5 mm occurred in elbows that contained none or nominally injured UCLs. In contrast ga pping greater than 0.5 mm was observed in elbows with large or complete te ars of the UCL. (Nassab, 2006) Figure 15: Imaging studies. (A) Plain anterpos terior radiograph of nonstressed elbow. (B) Stress anterposterior view of the same elbow Note the increase in space along the medial ulnohumeral joint line (arrow) (C) MRI depicting the cap sular “T” sign pathognomonic of a partial UCL tear (a rrow). (Nassab, 2006)

PAGE 41

30 A central objective to this research st udy was the assessment of gap formation of the reconstructions when subjected to a cycl ic valgus load at a 30 angle of flexion. Cyclic loading can assess graft slippage wh ich is imperative in the evaluation of early motion therapy protocols. In a clinical se tting, failure of a UCL reconstruction is generally attributed to slippage with resultan t laxity as opposed to traumatic graft rupture. (McAdams, 2007) 2.5.3 Ultimate Strength Research by Fleisig et al. has estimate d that the UCL resist s moments of 35 Nm during pitching and Werner et al. calculated th e actual forces to be 290 N. In previous work the intact cadaveric UCL was susceptible to a mo ment of 22.7 Nm under valgus stress. (Ahmad, 2003) A goal of this research study was to evaluate the ultimate strength and stiffness of the UCL reconstruction procedures. The results of this study will provide insight into an appropriate surgical method that mimics the anatomy and functionali ty of medial elbow stability.

PAGE 42

31 CHAPTER 3: METHODOLOGY 3.1 Overview Chapter 3 discusses the methodology and data analyzation of this research study. The first section details the instrumentati on used in this study followed by a section discussing specimen preparation. The ensuing sections describe the reconstructions performed and the te sting methodology. 3.2 Instrumentation In order to measure the biomechanical properties of the reconstructions two instruments were used during data collecti on and analyzation: a Test Resources 800L series (Shakopee, MN) servocontrolled ma terials testing machine and a Polhemus FASTRAK 3D motion tracking syst em (Colchester, VT). Both pieces of equipment were located at the Florida Orthopaedic Institute Biomechanics Laboratory (Tampa, FL). The following subsections detail the operative purpose and use of each instrument to this research study. 3.2.1 Servocontrolled Materials Testing Machine The purpose of this study was to compare the initial strength and biomechanical properties of three different UCL reconstruc tions with each other. To accomplish this purpose a cyclic valgus load was applied to the forearm constrained in supination, and

PAGE 43

32 distal to the anatomic axis of rotation of the elbow joint causing varus displacement. To achieve this purpose a servocontrolled test ing machine was employed that can measure biomechanical parameters of the elbow comp lex with the application of loads in both static and cyclic configurations. Performing a load or positi on controlled fatigue test re quires a servocontrolled test machine. Historically fatigue (also know n as tensile) testing machines have been servohydraulic or servocontrol of a hydraulic actuator. Fatigue tests are performed to select materials and to ensure quality for engineering appli cations. Of primary concern is material strength. Strength may be measured as stress necessary to cause plastic deformation or the maximum stress the material can withstand. Fatigue test are the most common material strength test and measure properties such as yield strength, modulus, ultimate tensile strength, Poisson’s ratio, re duction of area and elongation to failure. Additionally, tensile properties can predict material behavior under forms of loading other than tension. (Test Resources, WEB) The load frame is the basic structure of the standalone load unit of a materials testing machine. Two columns allow a cro sshead to be moved up or down to contain different size specimens and fixtures. The cr osshead and the base of the load frame are the two reaction masses in the force train. The linear actuator is mounted to the crosshead. It is a servocontrolled piston that applies displacement of (or force into) a specimen. It can apply equal power in te nsion and compression. One end of the test specimen is installed into a fixture mounted to the end of the actuator rod. The ax ial load unit includes a force transducer and a linear variable diffe rential transducer (LVDT) to measure linear forces and displacements. The force transduc er (also called load cell or force sensor) measures the amount of tension or compression a pplied to it. Preliminary testing was performed on Ar threx Fiberwire #5 a nd Fiberwire #2 to determine which suture had tensile propert ies similar to the palmaris longus tendon. During a test, a custom fixture is pin connected to the actuator to allow the application of a pure moment to the elbow complex rather than axial loading on the forearm. The pin connection allows transmission of a load to a specific location by ma intaining a constant

PAGE 44

33 lever arm. The displacements of and the for ces applied to the forearm are recorded via Test Resources data acquisition software on personal computer A (Dell Dimension 9200). Figure 16: Test Resources 800L series servoc ontrolled materials te sting machine. (Test Resources, WEB)

PAGE 45

34 Figure 17: Diagram of testing apparatus used in study. The actuator applies an upward valgus load (arrow) to potted radius and ulna (U). The po tted humerus (H) is fixed by a clamp. The setup allows for fixation (A) and pistoning (B) to maintain a constant lever arm length. M, medial; L, lateral. (McAdams, 2007) 3.2.2 3D Motion Tracking System Another objective of this study was to compare the elongations of the Jobe, Docking, and a novel Double bundle UCL recons truction procedure when subjected to cyclic valgus loading at 30 of flexion. To achieve this a Polhemus 3SPACE FASTRAK 3D motion tracking system was used. Through a built in digitize function dynamic real time measurements of position (X, Y, and Z Cartesian Coordinate s) and orientation (azimuth, elevation, and roll) can be recorde d. Displacement of the forearm results in the displacement of the associated digitized points which record s the elongation of the UCL reconstructions. The motion tracking system consists of a systems electronic unit (SEU), a power supply, one receiver, and one transmitter. The SEU encloses the hardware and software necessary to generate and se nse the magnetic fields, calcul ate position and orientation,

PAGE 46

35 and interface with the host computer. The transmitter contains electromagnetic coils enclosed in a plastic shell that emit the ma gnetic fields. The transmitter is the system’s reference frame for receiver measurements. Th e receiver, a lightweight cube contains electromagnetic coils enclosed in a molded pl astic shell that detect the magnetic fields emitted by the transmitter. The shape and weight of the receiver allows precise measurement of the receiver’s position a nd orientation. The recei ver is completely passive, having no active voltage ap plied to it. (Polhemus, WEB) During a test, a cyclic valgus load is a pplied to the medial elbow complex fixed in 30 of flexion. The displacements and the forc es applied to the forearm are recorded via data acquisition software on personal com puter A while the elongations of the UCL reconstructions are being track ed by the motion analysis sy stem and recorded on personal computer B (HP Compaq nc 8230). Figure 18: 3D electromagnetic motion tracking system.

PAGE 47

36 3.3 Specimen Preparation A 5 cm loop of Arthrex Fiberwire#2 a nd Fiberwire#5 using a Surgeon’s knot squared five times was prepared to test the knot breaking strength of each suture. The Surgeon’s knot was chosen for its popul arity among surgeons and its strength. Dr. Nofsinger, a committee member and attending surgeon at University Community Hospital in Tampa, Florida tied all of the Surgeon’s knots. A 5 cm loop was chosen to approximate the amount graft used in the re constructions. Each c onstruct was preloaded at 20N before each trial. The displaceme nt rate was 0.05 Hz. This experiment was repeated 5 times for each suture type. Looped suture specimen te sting eliminated Arthrex Fi berwire #2 as a simulation material for the palmaris longus graft. A st raight pull test was c onducted on Fiberwire#5 to determine the peak load of the suture ma terial. Each specimen was preloaded at 20N before each trial. A 14 cm suture specimen wa s attached to a custom fixture and pulled at a strain rate of 0.05 Hz. This expe riment was repeated 3 times. Figure 19: Straight pull test of Arthrex Fiberwire #5.

PAGE 48

37 Nine foam cortical shell elbows were obtained from Sawbones (Seattle, WA). Each elbow was osteomized 14 cm distal and pr oximal to the elbow joint. External latex bands were removed to facilitate the potting of the humerus and the forearm respectively. The bone ends were potted in 10 cm long a nd 1.5 in (approximately 3.8 cm) diameter cylindrical poly vinyl chloride (PVC) pi pe with body filler (Bondo). The forearm was fixed in supination to prevent stab ilization of the reconstruction. Figure 20: Disarticulated Sawbones with the humerus and forearms potted in PVC tubes with Bondo.

PAGE 49

38 3.4 Reconstructive Surgeries The procedures for reconstructive surgerie s used in this study were discussed in sections 2.3.2, 2.3.3 and 2.3.4. Each surgical pr ocedure was performed 3 times on a new sawbones specimen. The reconstructive su rgeries were performed by Dr. Charles Nofsinger a committee member and attending surgeon at University Community Hospital in Tampa, Florida. 3.5 Testing Methodology This study evaluates the biomechanics of three UCL reconstruction surgical procedures. All testing began with the posit ioning of each specimen in the materials testing machine. A complete test consiste d of one reconstructed elbow specimen and measured: valgus stability, gap fo rmation, and ultimate strength. Following the preparation of a specime n, discussed on section 3.3 its positioning to begin testing occurred. Specimens were fixed in 30 of flexion with the medial side of the elbow joint oriented supe riorly. A universal goniometer was used to measure 30 of flexion. The axis of the goniometer was placed just distal to the lateral epicondyle and both its proximal and distal arms were pl aced along the midline of the forearm and humerus. (Ellenbecker, 1998) The forearm was fi xed in the materials testing machine via a custom fixture. The potted humerus was fixed in an industrial vice. In order to measure the gap formati on of the reconstructions, the 3D motion tracking system was employed. The system track ed and calculated the gap formation of each reconstruction by the receiver that was zip-tied to the forearm near the reconstruction. For each test, the materials testing machine applied a moment to the potted forearm displacing the elbow in valgus. Duri ng the test, valgus stability, gap formation, and ultimate strength data were recorded and stored on personal computers A and B. A 20N preload was applied before each elbow joint was loaded to failure at 30 of flexion.

PAGE 50

39 Failure was defined as increase of gap forma tion of 5 mm or greater. The specimens were cyclically loaded and unloaded in valgus-varus for 200 cycles at 0.5 Hz at each load step beginning with 30N. If the repair survived 200 cycles, the cyclic load was increased by 10N; this was repeated in a stepwise fash ion until failure was achieved. (Pichora, 2007) The ultimate moment and length at instant of failure were recorded on personal computers A and B. The aforementioned steps ar e represented in a flow chart in Figure 21. Figure 21: Organizational flow char t depicting the testing methodology. Docking Tec hnique Jobe Technique Double Bundle Technique Applied Moment 30 of flexion Load displacement curves Gap Formation Valgus stability Methodology Flow Chart Specimens Input Test Output

PAGE 51

40 3.6 Data Consolidation The materials testing machine and the 3D motion tracking system were synchronized to acquire all data in real time. The materials testing machine recorded the applied load, the number of cycles and ultimate load to failure, while the 3D motion tracking system recorded the gap fo rmation of the reconstruction. Figure 22: Organizational flow char t depicting data consolidation. All data was consolidated into one set. The materials testing machine recorded date points every 0.08 seconds. The data acq uisition and digital servo-loop update occurred at a rate of 15-20 kHz. Sigma-de lta converters used to convert analog input signals have a data latency of up to 5-7 convers ion cycles to ensure in tegrity of the digital data fitting process. These latencies do not a ffect the quality of data processing or test Acquired Data Materials Testing Machine Computer A 3D Motion Analysis System Computer B Applied Moment Gap Formation Number of Cycles Load/Displacement Curves Data Consolidation

PAGE 52

41 data correlation because each input channel is independent. Resolution of dynamic control is of the order of 0.5% of required cyclic amplitude, or one point in 2500 of the designated active range (Test Resources, WEB). The 3D motion tracking system recorded data points every 0.02 seconds w ith a latency of 0.004 seconds. The static accuracy was 0.03 inches for Cartesian coordinates and 0.15 accuracy for orientation. The resolution is 0.0002 inches per inch of receiver and transmitter separation and 0.025 for orientation. (Polhemus, WEB) Valgus stability, gap formation and ultim ate strength were measured at cycles 50, 100, 200 and 600 or the cycle where failure oc curred. The aforementioned cycles were isolated from each respective data set and the mean displacement and load applied were isolated to calculate valgus stabilit y, gap formation, and ultimate strength.

PAGE 53

42 3.7 Processing and Analysis After data consolidation the test data was subsequently processed and analyzed. The results of this research were subdivid ed into valgus stability, gap formation, and ultimate strength for each of the three reconstruction procedures. The following subsections discuss the methodology for the pr ocessing and analyzati on of the data for each results section. 3.7.1 Valgus Stability In this research study valgus stability of the elbow joint was calculated by the change in valgus angle. This was accomp lished by measuring the displacement of the reconstruction correlated to a cyc lic valgus moment. This techni que is often referred to as the flexibility approach. From the data cons olidation the point with maximal load was isolated and that displacement was used. The moment arm of the valgus load wa s constantly applied at 140 mm from the joint line. Therefore, the displacement a ngle in radians was calculated from the arctangent of the ratio of the displaceme nt over the moment arm. This value is subsequently converted to degrees. (Shah, 2007) tan-1 = displacement X 180 Moment arm Figure 23: Valgus angle equation.

PAGE 54

43 3.7.2 Gap Formation In this research study the gap formation of the reconstruction procedures were measured as a function of flexion angle and valgus displacement due to a cyclic valgus moment. Prior to the cyclic loading of the reconstructions, the 3D motion tracking system recorded the initial position of the reconstr uctions. The measurement of the final position corresponds to the maximum applied valgus mo ment before gross re construction failure or an increase in length of 5 mm or greater. 3.7.3 Ultimate Strength The ultimate strength is a useful parameter in the evaluation of the overall stiffness and effectiveness of the reconstruction’ s ability to restore va lgus stability to the elbow joint. The elbow specimens were cyclically loaded to failure at 30 of flexion. The test began with the application of a cyc lic valgus load for 200 cycles until gross reconstruction failure or an incr ease in length of 5 mm or greater. The load was increased in a stepwise fashion for reconstructions that did not reach the modes of failure at the end of 200 cycles at the previous load step. The load at which failure occurred was recorded along with its associat ed displacement. Each test yielded a load to failure curve in which the failure value is depicted as an abrupt change in torque at the moment of failure. The ultimate strength value calculated is the ratio of the ultimate load to its associated displacement or the slope of the load as a function of displacement graph. Moment = Newton meters Figure 24: Moment calculation.

PAGE 55

44 CHAPTER 4: RESULTS 4.1 Overview Chapter 4 discusses the results of the proposed research study and is subdivided into five major sections which includ e the results of: tendon simulations, Jobe reconstruction, Docking reconstruction, Doubl e bundle reconstruction and a comparative analysis of all three reconstr uction procedures. Each of the reconstruction results sections is further subdivided into: valgus stabili ty, gap formation and ultimate strength. 4.2 Tendon Simulations The knot load to failure and the elonga tion of each looped suture specimen is presented in Table 1. The stiffness of the c onstructs was defined as the linear portion of the stress-strain curv e and is shown in Table 1 and Figure 25. The ultimate load to failure and corre sponding elongation of each suture specimen is presented in Table 2. The stiffness of th e suture specimens is shown in Figure 26.

PAGE 56

45 Table 1: Knot maximum elongation and maximu m load and stiffness values for Arthrex Fiberwire #2 and #5. Arthrex Fiberwire #5 Trial Maximum Elongation (mm) Knot Maximum Load (N) Stiffness (N/mm) 1 5.2 178.1 34.25 2 5.97 185 30.99 3 5.26 172.9 32.87 Arthrex Fiberwire #2 Trial Maximum Elongation (mm) Knot Maximum Load (N) Stiffness (N/mm) 1 3.95 102.9 26.05 2 3.27 74.6 22.81 3 5.14 85.5 19.57 4 5.16 88.9 19.41 Table 2: Recorded maximum load and ma nufacturer’s maximum load for Arthrex Fiberwire #2 and #5. Suture Material Recorded Maximum Load (N) Manufacturer's Maximum Load (N) Arthrex Fiberwire #2 236.68 271.5 Arthrex Fiberwire #5 400 600

PAGE 57

46 Stiffness (N/mm) of Arthrex Fiberwire #20 20 40 60 80 100 120 012345 Displacement (mm)Force (N) Series1 Series2 Series3 Series4 Stiffness (N/mm) of Arthrex Fiberwire #50 20 40 60 80 100 120 140 160 180 200 01234567 Displacement (mm)Force (N) Series1 Series2 Series3 Figure 25: Graphs depicting th e stiffness of looped suture specimens (A) Fiberwire #2 (B) Fiberwire #5. A B

PAGE 58

47 Arthrex Fiberwire #5 Suture Straight Pull Test0 50 100 150 200 250 300 350 400 0510152025 Displacement (mm)Force (N) Series1 Figure 26: Graphs depicting the stiffness of su ture specimens with a straight pull test (A) suture specimen 1 (B) suture specimen 2. Arthrex Fiberwire #5 Straight Pull Test0 50 100 150 200 250 300 350 400 02468101214 Displacement (mm)Force (N) Series1 A B

PAGE 59

48 4.3 Jobe Reconstructed Ulnar Collateral Ligament 4.3.1 Valgus Stability The individual and mean valgus stability and flexibility values for Jobe reconstructed specimens at 30 of flexi on for cycles 50, 100, 200 and 600 are displayed in Tables 3 and 4. The accompanying graphs of individual and mean valgus stability and flexibility as a function of cycles 50, 100, 200 and 600 are depicted in Figures 27 and 28 respectively. Table 3: Individual valgus st ability and flexibility for the Jobe reconstructed UCL at cycles 50, 100, 200 and 600. Cycle Specimen Valgus stability (deg) flexibility (deg/Nm) 50 1 0.42 0.25 100 1 0.42 0.25 200 1 0.31 0.19 600 1 0.45 0.23 50 2 0.34 0.09 100 2 0.33 0.08 200 2 0.33 0.08 600 2 1.06 0.25

PAGE 60

49 Table 4: Mean and standard deviation values of valgus stability and flexibility for the Jobe reconstructed UCL at cycles 50, 100, 200 and 600. Cycle valgus stability (deg) flexibility (deg/Nm) Mean 50 0.38 0.17 Std Deviation 0.05 0.11 Mean 100 0.37 0.17 Std Deviation 0.06 0.12 Mean 200 0.32 0.13 Std Deviation 0.02 0.08 Mean 600 0.76 0.24 Std Deviation 0.43 0.02

PAGE 61

50 Valgus Stability of Individual Jobe Reconstructions0 0.2 0.4 0.6 0.8 1 1.2 1.4 cycle 50cycle 100cycle 200cycle 600 CyclesValgus Stability (degrees) Series1 Series2 Mean Valgus Stability of Jobe Reconstructed UCL0.00 0.05 0.10 0.15 0.20 0.25 0.30 0.35 0.40 cycle 50cycle 100cycle 200cycle 600 CyclesValgus stability (degrees) Series1 Figure 27: Valgus stability of Jobe UCL rec onstructions for (A) indi vidual specimens and (B) and the mean of specimens. A B

PAGE 62

51 Flexibility of Individual Jobe Reconstructions0 0.05 0.1 0.15 0.2 0.25 0.3 0.35 cycle 50cycle 100cycle 200cycle 600 CyclesFlexibility (degrees/Nm) Series1 Series2 Mean Flexibility of Jobe Reconstructions0.00 0.05 0.10 0.15 0.20 0.25 0.30 0.35 cycle 50cycle 100cycle 200cycle 600 CyclesFlexibility (degrees/Nm) Series1 Figure 28: Flexibility of Jobe UCL reconstr uctions for (A) individual specimens and (B) the mean of specimens. A B

PAGE 63

52 4.3.2 Gap Formation The maximal displacement as a result of the application of a valgus moment was isolated for gap formation calculations. The in dividual and mean gap formation values of Jobe reconstructed specimens at 30 of flexion for cycles 50, 100, 200 and 600 are displayed in Tables 5 and 6. The accompa nying graphs of individual and mean gap formation as a function of cycles 50, 100, 200 and 600 are depicted in Figure 29. Table 5: Individual gap formation for the Jobe reconstructed UCL at cycles 50, 100, 200 and 600. Cycle Specimen Gap formation (mm) 50 1 1.02 100 1 1.02 200 1 0.76 209 1 1.11 50 2 0.84 100 2 0.81 200 2 0.81 250 2 2.60

PAGE 64

53 Table 6: Mean and standard deviation values of gap formation for the Jobe reconstructed UCL at cycles 50, 100, 200 and 600. Cycle Gap Formation (mm) Mean 50 0.93 Std Deviation 0.13 Mean 100 0.92 Std Deviation 0.14 Mean 200 0.79 Std Deviation 0.04 Mean 600 1.85 Std Deviation 1.06

PAGE 65

54 Gap Formation of Jobe Reconstructions0 0.5 1 1.5 2 2.5 3 3.5 cycle 50cycle 100cycle 200cycle 600 CycleGap formation (mm) Series1 Series2 Mean Gap Formation for Jobe Reconstructions0 0.5 1 1.5 2 2.5 cycle 50cycle 100cycle 200cycle 600 CycleGap formation (mm) Series1 Figure 29: Gap formation of Jobe UCL recons truction for (A) indivi dual specimens and (B) the mean of specimens. A B

PAGE 66

55 4.3.3 Ultimate Strength The ultimate strength required to either completely disrupt the Jobe reconstructed UCL or create a gap formation of 5 mm or greater is displayed in Table 7 for each specimen at cycles 50, 100, 200 and 600. Table 8 displays the mean ultimate strength required for failure of the Jobe reconstr uction at cycles 50, 100, 200 and 600. Graphical representation of both the individual and mean ultimate strength is shown in Figure 30-A and Figure 30-B respectively. Figure 31 graphi cally displays the number of cycles to failure for each Jobe reconstruction specimen. Table 7: Individual ultimate strength for the Jobe reconstructed UCL at cycles 50, 100, 200 and 600. Cycle Specimen Ultimate strength (Nm) 50 1 4.10 100 1 4.08 200 1 4.06 209 1 4.85 50 2 4.01 100 2 4.04 200 2 4.13 250 2 4.18

PAGE 67

56 Table 8: Mean and standard deviation values of ultimate strength for the Jobe reconstructed UCL at cycles 50, 100, 200 and 600. Cycle Ultimate Strength (Nm) Mean 50 4.06 Std Deviation 0.07 Mean 100 4.06 Std Deviation 0.03 Mean 200 4.09 Std Deviation 0.05 Mean 600 4.51 Std Deviation 0.47

PAGE 68

57 Ultimate Strength of Individual Jobe Reconstructions0 1 2 3 4 5 6 7 cycle 50cycle 100cycle 200cycle 600 CycleUltimate Strength (Nm) Series1 Series2 Mean Ultimate Strength of Jobe Reconstructions3.60 3.80 4.00 4.20 4.40 4.60 4.80 cycle 50cycle 100cycle 200cycle 600 CycleUltimate Strength (Nm) Series1 Figure 30: Ultimate strength of Jobe UCL reconstructions for (A) individual specimens and (B) the mean of specimens. A B

PAGE 69

58 Number of Cylces to Failure for Jobe Reconstructions0 50 100 150 200 250 300 specimen 1specimen 2 SpecimensCycles Series1 Figure 31: Number of cycles to failu re for each Jobe UCL reconstruction. 4.4 Docking Reconstructed Ulnar Collateral Ligament 4.4.1 Valgus Stability The individual and mean valgus stabili ty and flexibility values for Docking reconstructed specimens at 30 of flexi on for cycles 50, 100, 200 and 600 are displayed in Tables 9 and 10. The accompanying graphs of individual and mean valgus stability and flexibility as a function of cycles 50, 100, 200 and 600 are depicted in Figures 32 and 33.

PAGE 70

59 Table 9: Individual valgus stab ility and flexibility for the Docking reconstructed UCL at cycles 50, 100, 200 and 600. Cycle Specimen valgus stability (deg) flexibility (deg/Nm) 50 1 0.35 0.09 100 1 0.31 0.08 200 1 0.43 0.11 600 1 0.95 0.14 50 2 0.89 0.24 100 2 1.10 0.28 200 2 1.35 0.34 600 2 1.03 0.22 50 3 0.64 0.18 100 3 0.70 0.18 200 3 0.79 0.19 600 3 0.52 0.08 Table 10: Mean and standard deviation values of valgus stability and flexibility for the Docking reconstructed UCL at cycles 50, 100, 200 and 600. Cycle valgus stability (deg) flexibility (deg/Nm) Mean 50 0.63 0.17 Std Deviation 0.27 0.08 Mean 100 0.70 0.18 Std Deviation 0.40 0.10 Mean 200 0.86 0.21 Std Deviation 0.46 0.12 Mean 600 0.83 0.14 Std Deviation 0.27 0.05

PAGE 71

60 Valgus Stabilty for Individual Docking reconstructions0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 cycle 50cycle 100cycle 200cycle 600 CycleValgus Stability (degrees) Series1 Series2 Series3 Mean Valgus Stability of Docking Reconstructions0.00 0.20 0.40 0.60 0.80 1.00 1.20 cycle 50cycle 100cycle 200cycle 600 CycleValgus Stability (degrees) Series1 Figure 32: Valgus stability of Docking UCL reconstructions for (A) individual specimens and (B) the mean of specimens. A B

PAGE 72

61 Flexibility of Individual Docking Reconstructions0.00 0.05 0.10 0.15 0.20 0.25 0.30 0.35 0.40 0.45 cycle 50cycle 100cycle 200cycle 600 CycleFlexibility (degree/Nm) Series1 Series2 Series3 Mean Flexibility of Docking Reconstructions0.00 0.05 0.10 0.15 0.20 0.25 0.30 cycle 50cycle 100cycle 200cycle 600 CycleFlexibility (degrees/Nm) Series1 Figure 33: Flexibility of Docking UCL recons tructions for (A) individual specimens and (B) the mean of specimens. A B

PAGE 73

62 4.4.2 Gap Formation The maximal displacement as a result of the application of a valgus moment was isolated for gap formation calculations. The in dividual and mean gap formation values of Docking reconstructed specimens at 30 of flexion for cycles 50, 100, 200 and 600 are displayed in Tables 11 and 12. The accompa nying graphs of indi vidual and mean gap formation as a function of cycles 50, 100, 200 and 600 are depicted in Figure 34. Table 11: Individual gap formation for the Do cking reconstructed UCL at cycles 50, 100, 200 and 600. Cycle Specimen Gap Formation (mm) 50 1 0.85 100 1 0.75 200 1 1.06 600 1 2.32 50 2 2.18 100 2 2.69 200 2 3.30 600 2 2.52 50 3 1.57 100 3 1.72 200 3 1.93 600 3 1.27

PAGE 74

63 Table 12: Mean and standard deviation va lues of gap formation for the Docking reconstructed UCL at cycles 50, 100, 200 and 600. Cycle Gap Formation (mm) Mean 50 1.53 Std Deviation 0.67 Mean 100 1.72 Std Deviation 0.97 Mean 200 2.10 Std Deviation 1.13 Mean 600 2.04 Std Deviation 0.67

PAGE 75

64 Individual Gap Formation of Docking Reconstructions0 0.5 1 1.5 2 2.5 3 3.5 4 cycle 50cycle 100cycle 200cycle 600 CycleGap Formation (mm) Series1 Series2 Series3 Mean Gap Formation of Docking Reconstructions0.00 0.50 1.00 1.50 2.00 2.50 3.00 cycle 50cycle 100cycle 200cycle 600 CycleGap Formation (mm) Series1 Figure 34: Gap formation of Docking UCL r econstructions for (A) individual specimens and (B) the mean of specimens. A B

PAGE 76

65 4.4.3 Ultimate Strength The ultimate strength required to ei ther completely disrupt the Docking reconstructed UCL or create a gap formation of 5 mm or greater is displayed in Table 13 for each specimen at cycles 50, 100, 200 and 600. Table 14 displays the mean ultimate strength required for failure of the Dock ing reconstruction at cycles 50, 100, 200 and 600. Graphical representation of both the individua l and mean ultimate strength is shown in Figure 35-A and Figure 35-B resp ectively. Figure 36 graphica lly displays the number of cycles to failure for each Do cking reconstruction specimen. Table 13: Individual ultimate strength for the Docking reconstructed UCL at cycles 50, 100, 200 and 600. Cycle Specimen Ultimate strength (Nm) 50 1 3.96 100 1 3.97 200 1 4.01 600 1 6.89 50 2 3.68 100 2 3.87 200 2 3.99 600 2 4.68 50 3 3.61 100 3 3.96 200 3 4.06 600 3 6.89

PAGE 77

66 Table 14: Mean and standard deviation values of ultimate strength for the Docking reconstructed UCL at cycles 50, 100, 200 and 600. Cycle Ultimate Strength (Nm) Mean 50 3.75 Std Deviation 0.19 Mean 100 3.93 Std Deviation 0.06 Mean 200 4.02 Std Deviation 0.03 Mean 600 3.87 Std Deviation 1.28

PAGE 78

67 Individual Ultimate Strength of Docking Reconstructions0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 cycle 50cycle 100cycle 200cycle 600 CycleUltimate Strength (Nm) Series1 Series2 Series3 Mean Ultimate Strength of Docking Reconstructions3.50 3.60 3.70 3.80 3.90 4.00 4.10 cycle 50cycle 100cycle 200cycle 600 CycleUltimate Strength (Nm) Series1 Figure 35: Ultimate strength of Docking UCL reconstructions for (A) individual specimens and (B) the mean of specimens. A B

PAGE 79

68 Number of Cycles to Failure0 100 200 300 400 500 600 700 800 specimen 1specimen 2specimen 3 SpecimenCycle Series1 Figure 36: Number of cycles to failu re for each Docking reconstruction. 4.5 Double Bundle Reconstructed Ulnar Collateral Ligament 4.5.1 Valgus Stability The individual and mean valgus stabili ty and flexibility values for Double bundle reconstructed specimens at 30 of flexi on for cycles 50, 100, 200 and 600 are displayed in Tables 15 and 16. The accompanying graphs of individual and mean valgus stability and flexibility as a function of cycles 50, 100, 200 and 600 are depicted in Figures 37 and 38 respectively.

PAGE 80

69 Table 15: Individual valgus st ability and flexibility for the Double bundle reconstructed UCL at cycles 50, 100, 200 and 600. Cycle Specimen valgus stability (deg) flexibility (deg/Nm) 50 1 0.49 0.12 100 1 0.50 0.12 200 1 0.51 0.13 600 1 0.30 0.05 50 2 0.62 0.16 100 2 0.50 0.18 200 2 0.70 0.18 600 2 0.48 0.09 Table 16: Mean and standard deviation values of valgus stability and flexibility for the Double bundle reconstructed UCL at cycles 50, 100, 200 and 600. Cycle valgus stability (deg) flexibility (deg/Nm) Mean 50 0.55 0.14 Std Deviation 0.09 0.02 Mean 100 0.60 0.15 Std Deviation 0.15 0.04 Mean 200 0.61 0.15 Std Deviation 0.14 0.03 Mean 600 0.39 0.07 Std Deviation 0.13 0.02

PAGE 81

70 Individual Valgus Stability of Double Bundle Reconstructions0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 cycle 50cycle 100cycle 200cycle 600 CycleValgus Stability (degrees) Series1 Series2 Mean Valgus Stability of Double Bundle Reconstructions0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 cycle 50cycle 100cycle 200cycle 600 CycleValgus Stability (degrees) Series1 Figure 37: Valgus stability of Double bundl e UCL reconstructions for (A) individual specimens and (B) the mean of specimens. A B

PAGE 82

71 Individual Flexibility of Doube Bundle Reconstructions0.00 0.05 0.10 0.15 0.20 0.25 cycle 50cycle 100cycle 200cycle 600 CycleFlexibility (degrees/Nm) Series1 Series2 Mean Flexibility of Double Bundle Reconstructions0.00 0.02 0.04 0.06 0.08 0.10 0.12 0.14 0.16 0.18 0.20 cycle 50cycle 100cycle 200cycle 600 CycleFlexibility (degrees/Nm) Series1 Figure 38: Flexibility of Double bundle UC L reconstructions for (A) individual specimens and (B) the mean of specimens. A B

PAGE 83

72 4.5.2 Gap Formation The maximal displacement as a result of the application of a valgus moment was isolated for gap formation calculations. The in dividual and mean gap formation values of Double bundle reconstructed specimens at 30 of flexion for cycles 50, 100, 200 and 600 are displayed in Tables 17 and 18. The accompa nying graphs of individual and mean gap formation as a function of cycles 50, 100, 200 and 600 are depicted in Figure 39. Table 17: Individual gap formation for the D ouble bundle reconstructed UCL at cycles 50, 100, 200 and 600. Cycle Specimen Gap Formation (mm) 50 1 1.19 100 1 1.21 200 1 1.25 600 1 0.73 50 2 1.51 100 2 1.73 200 2 1.72 600 2 1.17

PAGE 84

73 Table 18: Mean and standard deviation va lues of gap formation for the Double bundle reconstructed UCL at cycles 50, 100, 200 and 600. Cycle Gap Formation (mm) Mean 50 1.35 Std Deviation 0.23 Mean 100 1.47 Std Deviation 0.36 Mean 200 1.49 Std Deviation 0.33 Mean 600 0.95 Std Deviation 0.31

PAGE 85

74 Individual Gap Formation of Double Bundle Reconstructions0.00 0.50 1.00 1.50 2.00 2.50 cycle 50cycle 100cycle 200cycle 600 CycleGap Formation (mm) Series1 Series2 Mean Gap Formation of Double Bundle Reconstructions0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 cycle 50cycle 100cycle 200cycle 600 CycleGap Formation (mm) Series1 Figure 39: Gap formation of Double bundle UCL reconstructions for (A) individual specimens and (B) the mean of specimens. A B

PAGE 86

75 4.5.3 Ultimate Strength The ultimate strength required to eith er completely disrupt the Double bundle reconstructed UCL or create a gap formation of 5 mm or greater is displayed in Table 19 for each specimen at cycles 50, 100, 200 and 600. Table 20 displays the mean ultimate strength required for failure of the D ouble bundle reconstruction at cycles 50, 100, 200 and 600. Graphical representation of both the individual and mean ultimate strength is shown in Figure 40-A and Figure 40-B respec tively. Figure 41 graphically displays the number of cycles to failure for each Double bundle reconstruction specimen. Table 19: Individual ultimate strength for th e Double bundle reconstructed UCL at cycles 50, 100, 200 and 600. Cycle Specimen strength (Nm) 50 1 3.96 100 1 4.06 200 1 3.96 600 1 5.44 50 2 3.96 100 2 4.04 200 2 4.01 600 2 5.33

PAGE 87

76 Table 20: Mean and standard deviation valu es of ultimate strength for the Double bundle reconstructed UCL at cycles 50, 100, 200 and 600. Cycle strength (Nm) Mean 50 3.96 Std Deviation 0.00 Mean 100 4.05 Std Deviation 0.01 Mean 200 3.98 Std Deviation 0.03 Mean 600 5.39 Std Deviation 0.04

PAGE 88

77 Individual Ultimate Strength of Double Bundle Reconstructions0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 cycle 50cycle 100cycle 200cycle 600 CycleUltimate Strength (Nm) Series1 Series2 Mean Ultimate Strength of Double Bundle Reconstructions0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 cycle 50cycle 100cycle 200cycle 600 CycleUltimate Strength (Nm) Series1 Figure 40: Ultimate strength of Double bundle UCL reconstructions for (A) individual specimens and (B) the mean of specimens. A B

PAGE 89

78 Number of Cycles to Failure401 401.5 402 402.5 403 403.5 specimen 1specimen 2 SpecimenCycles Series1 Figure 41: Number of cycles to failu re for each Double bundle reconstruction.

PAGE 90

79 CHAPTER 5: DISCUSSION 5.1 Overview Chapter 5 discusses the results of the previous chapter. The preceding section compares the results of the tendon simulation ex periments with earlie r studies that have documented the mechanical properties of the palmaris longus. A subsequent analysis of the capabilities of th e Jobe, Docking, and Double bundle reconstruction procedures to restore the function of the intact UCL will ensue. The analysis is subdivided by the biomechanical parameters measured--valgus stability, gap formation, and ultimate strength. The functionality of each reconstruction will be assessed according to the functionality of the intact UCL as published in ear lier studies. 5.2 Tendon Simulations The ipsilateral palmaris longus tendo n is the primary graft source for UCL reconstructions. (Prud’homme, 2008) Additio nal graft choices include: gracilis, semitendinosous, and patellar tendon grafts Three independent research studies Prud’homme et al., Simonian et al., and Wilson et al have shown no significant difference in the load to failure, stiffness, or modulus of elasticity of UCL reconstructions with the aforementioned graft sources. Therefor e, the graft source most readily available with the lowest association of morbidity should be used. (Prud’homme, 2008) This graft source is usually the palmaris longus. Resear ch by Prud’homme et al. showed that the elasticity of the tendon is the limiting factor in ligament reconstructions. As a result, all tendon simulation data was compared with the stiffness of the palmaris longus.

PAGE 91

80 The current study compared the mechanic al properties of Arthrex Fiberwire #2 and Fiberwire #5 with the palmaris longus tendo n. The load to failure was defined as the amount of force (N) necessary to cause catas trophic rupture or knot breakage of the specimen. Stiffness was defined as the force required to deform a unit length. There was no significant difference in the load to failu re of Fiberwire #2 in comparison with the manufacturer’s data (Table 2) Although there was a significant difference in the load to failure of the Arthrex Fiberw ire #5 and the manufacturer’s data, there was no significant difference in the stiffness measurement. Th e stiffness property of Arthrex Fiberwire #5 most closely resembled the stiffness pr operties of the palmaris longus and was subsequently used as a simulated ten don in the reconstruction experiments. Regan et al. determined the ultimate lo ad of the palmaris longus tendon to be 358N. Prud’homme et al. quantif ied the stiffness of the pa lmaris longus tendon to be 30.78 N/mm with a standard deviation of 31.57. Th is data is statisti cally similar to the stiffness of Arthrex Fiberwire #5 which serv ed as the rationale for its use as tendon substitute. The differences in the load to failure da ta of the manufacturer and this research can likely be attributed to differences in loading protoc ol. The manufacturer’s test parameters utilized custom jig s with ambient air at a rate of 0.08 Hz, whereas this study did not utilize suture or cord specific grips. Furthermore the loading rate was slightly faster at 0.05 Hz. An additional area of con cern was that of knot settling and slipping which causes a non-recoverable displacement of the fixation. A possible solution to the knot settling problem would be to position the humerus and the ulna at an increased angle of flexion and then secure the fixation. As a result, when the knot settles the humerus and the ulna will be positione d correctly relative to ea ch other. (Harrell, 2003)

PAGE 92

81 5.3 Comparison of Surgical Reeconstructions The most advantageous method of re construction of the UCL has received minimal focus in the literature. This research has shown that each of these reconstructions failed at significantly lower loads than the reported values of the intact UCL. Additionally, the peak loads of this research were significantly lower than those reported in the previous research. One explanation fo r this difference may be the cyclic loading protocol used in this researc h, whereas the majority of other studies used a load-to-failure protocol. Load-to-failure and cyclic loading are the two types of loading protocols that are used to assess the fixation of soft tissue. Load-to-fail ure protocols only measure the ability of the ligament to with stand traumatic rupture. A cyc lic loading protocol is more meaningful as it mimics early postoperative rehabilitation. Litera ture on flexor tendon repairs suggests that a cyclic loading prot ocol will weaken the repair by producing gap formation and subsequently lead to failure at lower loads than a lo ad-to-failure protocol. There was no complete disruption of any of the reconstructions. The mode of failure for each was gap formation of the reconstruc tion. After cyclic loading, there was a macroscopically visible groove in the Sawbone from the si mulated tendon which resulted in macroscopic bending of the head of the humerus during cyclic loading and may have affected the results. To compare the reconstructions across speci fic cycles a statistical analysis of the data was performed with the use of the SAS statistical software package (Cary, NC). The valgus stability, gap formation and ultimat e strength for the Jobe, Docking, and Double bundle techniques were compared at cycles 50, 100, 200 and 600 by the use of a repeated measures analysis of variance (ANOVA) with a significance criterion of 0.05. Differences between groups were analyzed wi th Tukey’s Honestly Significant Difference test.

PAGE 93

82 5.3.1 Valgus Stability Figure 42 compares the mean valgus stab ility for the Jobe, Docking, and Double bundle reconstructions at cycles 50, 100, 200 and 600. A significance level of 0.05 showed no differences between any of the reco nstructions for any of the cycles measured. The low standards of deviation in the values suggest consistency in the testing protocol. One specimen in the Double bundle group stre tched tremendously unde r initial loading, resulting in a large variability in displ acement in peak load for this group which contributed to the larger y-bar error in th is group. None of the separate univariate ANOVAs was statistically significant. There was no statistically significant difference; however the power is not high enough to state wi th confidence that st atistically there is no difference. Valgus Stability Comparison0 0.2 0.4 0.6 0.8 1 1.2 cycle 50cycle 100cycle 200cycle 600 CycleValgus Stability (degrees) Jobe Docking Double bundle Figure 42: Valgus stability comparison of the Jobe, Docking and Double bundle reconstructions at cycles 50, 100, 200 and 600.

PAGE 94

83 5.3.2 Gap Formation In each reconstruction group, gap formation was observed as soon as a load was applied. This early gap formation occurred despite a 20N preload applied to each specimen. In the Jobe and Docking reconstructions the mean displacement after application of a 30N load was 2.63 mm and 2.10 mm respectively. According to the research of Ellenbecker et al. the Jobe rec onstruction has already reached clinical failure. Table 21 represents the mean displacement at each load step for each reconstruction. A significance level of 0.05 showed no differences between any of the reconstructions for any of the cycles meas ured. None of the separate univariate ANOVAs was statistically significant. There was no sta tistically significant difference; however the power is not high enough to state with confidence that statistically th ere is no difference. Table 21: Depicts the mean displacement at each load step for the Jobe, Docking, and Double bundle reconstructions. Jobe Docking Double bundle Load (N) Mean Displacement (mm) No. of Specimens Mean Displacement (mm) No. of Specimens Mean Displacement (mm) No. of Specimens 30 2.63 0.23 2 2.10 1.13 3 1.38 0.50 2 40 5.19 0.18 2 4.07 1.62 3 2.915 0.22 2 50 NA 0 4.97 1 5.24 0.59 2

PAGE 95

84 Comparison of Gap Formation0.00 0.50 1.00 1.50 2.00 2.50 3.00 cycle 50cycle 100cycle 200cycle 600 CycleGap Formation (mm) Jobe Docking Double bundle Figure 43: Gap formation comparison of the Jobe, Docking and Double bundle reconstructions at cycles 50, 100, 200 and 600. 5.3.3 Ultimate Strength The ultimate strength and peak loads reached by each reconstructed specimen are significantly lower than the previously reporte d values for the intact UCL. Armstrong et al. measured a mean peak load to failure of 142.5N. In the present study the Jobe reconstruction obtained a mean peak load a nd cycles to failure of 34.62N and 229.5cycles, the Docking reconstruction 49.24N a nd 411.33 cycles and the Double bundle reconstruction 38.85N and 402.5 cycles respective ly. These results are consistent with those of Armstrong et al. None of the sepa rate univariate ANOVAs was statistically significant. There was no statistically signi ficant difference; however the power is not high enough to state with confidence that statistically ther e is no difference. Figure 44 compares the mean ultimate strength for the Jobe, Docking, and Double bundle reconstructions at cycles 50, 100, 200 and 600. A significance level of 0.05 showed no differences between any of the reco nstructions for any of the cycles measured.

PAGE 96

85 Comparison of Ultimate Strength0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 cycle 50cycle 100cycle 200cycle 600 CycleUltimate Strength Jobe Docking Double bundle Figure 44: Ultimate strength comparison of the Jobe, Docking and Double bundle reconstructions at cycles 50, 100, 200 and 600. Mean Number of Cycles to Failure0 50 100 150 200 250 300 350 400 450 500 JobeDockingDouble bundle ReconstructionsCycles Series1 Figure 45: Mean number of cycles to failu re of the Jobe, Docking, and Double bundle reconstructions.

PAGE 97

86 CHAPTER 6: SUMMARY The elbow is a synovial hinge joint that is composed of various skeletal and soft tissue constraints that contribu te to its anatomy and functi on. Skeletal anatomy such as the olecranon/olecrenon fossa ar ticulation provide primary stability of less than 20 or greater than 120 of elbow flexion. Soft tissue constraints provide the elbow with dynamic and static stability during the overhead arc of motion (20-120 ). There are three primary functions of the elbow: to serves as a component joint of the lever arm in positioning the hand, a fulcrum for the forearm lever, and a weight bearing joint in patients using crutches (Miller, 1992) The anterior band (AB) is taut during the first 60 of flexion and the posterior band (PB) is taut between 60 and 120 of flex ion. This provides a reciprocal function in resisting valgus stress duri ng flexion and extension. The central band is isometric throughout flexion and extens ion. (Safran, 2005, Nassab, 2006) In this research study the valgus stability of the reconstructe d elbow as a function of cyclic valgus loading was evaluated. A stronger UCL reconstr uction should reduce the valgus angle in response to the application of a valgus moment A central objective to this research study was the assessment of gap formation of the reconstructions when subjected to a cyclic valgus lo ad at a 30 angle of flexion. Cy clic loading can assess graft slippage which is imperative in the evaluati on of early motion therapy protocols. In a clinical setting, failure of a UCL reconstruction is generally attributed to slippage with resultant laxity as opposed to trau matic graft rupture. (McAdams, 2007) This research compared the valgus stability, gap formation, and ultimate strength that resulted from the cyclic valgus loading of 3 techniques for reconstruction of the UCL: the Jobe, Docking, and a novel Double bundle pr ocedure. A servocont rolled materials

PAGE 98

87 testing machine applied a cyclic valgus load to white cortical Sawbones elbow complex models while a 3D motion track ing system recorded the valg us displacement of the UCL reconstructions. The valgus stability, gap formation, and ultimate strength were measured at 50, 100, 200 and 600 cycles or the cycle at which fa ilure occurred. The mean peak load to failure was 30N for the Jobe reconstructi ons and 50N for both the Docking and Double bundle reconstructions. Both the Docki ng and the Double bundle reconstructions sustained a higher load to failure than the Jobe reconstruction. None of the separate univariate ANOVAs of the biomechanical parameters of each reconstruction was statistically significant. Alt hough there was no statistically significant difference, a small standard deviation in all measured values indicated consistenc y in testing methodology. The power or sample size is not high enough to state with confidence that statistically there is no difference.

PAGE 99

88 CHAPTER 7: RECOMMENDATIONS FOR FUTURE WORK This research study was completed with no major complications. Both the servocontrolled materials testing machine a nd the 3D motion analys is system functioned with ease throughout the experi mentation. Although subsequent data analysis produced consistent results in the in vitro protocol, the use of s ynthetic elbows (Sawbones) produced relative values for the biomechanical parameters that do not correlate with cadaveric values. The ensuing paragraphs detail improvements for new studies which include: the use of composite bone material s, cadaver specimens, kinematic testing, in vitro testing and surgeon variability. Biomechanical values can be obtained w ith the use of composite bone materials or cadaver specimens. Composite bones are desi gned to simulate the physical behavior of a human bone, providing an alternative fo r cadaver bones in testing and research. Mechanical behavior of the composite bone ma terial falls within the range for cadaveric specimens. Composite bones have shown signi ficantly lower variability in testing compared to cadaveric specimens for all load ing regimens, offering a more reliable test bed. Other advantages of testing with co mposite bones include unlimited sample sizes with no special handling or preservative requirements. (Sawbones, WEB) In order to effectively treat elbow instability, a thorough understanding of the contributions of both dynamic and static st ructures must be understood. (Safran, 2005) The effect of the degree of soft tissue di ssection can only be measured in cadaver specimens. One could assume that less soft tissue dissection woul d improve the overall strength of reconstruction. (Hechtman, 1998). This was not a kinematic study, this study investigated the response of a cyclic valgus load at a fixed angle rather than throughout the ROM. Evaluation other than at time=0 will te st the effect of active muscle contraction

PAGE 100

89 or joint compression. Testing at various angles of elbow flexion would better approximate a clinical situation. (Pale tta, 2006) Additionally, the repair strength evaluated at time=0 does not examine the effect of early postoperati ve tissue healing on valgus stability. (Pichora, 2007) Further study is needed to determin e the correlation with the number of cycles with rehabilitation pr ogram as graft healing occurs. (McAdams, 2007) Accurate replication of the flexion-exte nsion axis of the elbow is important when using articulated external fixators and when performing ligamentous reconstructions. To study the effects of surgeon va riability in tunnel placement on the biomechanics of UCL reconstructions, the anchor points of the su rgery could be varied according the surgeon repeatability data obtained from Brownhill et al. Correct selection of the flexionextension axis allows for pr oper placement of the implan t, thereby recreating normal biomechanics, including the natural joint motion and muscle moment arms about the elbow. This is in all likelihood, important with regard to pr eserving natural joint forces and kinematics.

PAGE 101

90 REFERENCES "Biomechanical Test Materials." Re trieved March, 2008, from sawbones.com "Dynamic Test Systems: 800L seri es." Retrieved March, 2008, from http://www.testresources.com/systems/800L.aspx. "Polhemus Innovation in Motion: 3SPACE FASTRAK." Retrieved March, 2008, from http://www.polhemus.com/?page=Motion_Fastrak. (1997). "Musculoskeletal Radiology” Prona tor Teres Retrieved March, 2008, from http://www.rad.washington.edu/ academics/academic-sections /msk/muscle-atlas/upperbody/pronator-teres (2000). "Gray's Anatomy of the Human Body." Elbow Joint Retrieved March, 2008, from http://education.yahoo.com /reference/gray/s ubjects/subject/84. Ahmad, C. S., T. Q. Lee, et al. (2003). "Biomechanical evaluation of a new ulnar collateral ligament reconstruc tion technique with interfer ence screw fixation." Am J Sports Med 31(3): 332-7. Armstrong, A. D., C. E. Dunning, et al. (2005) "A biomechanical comparison of four reconstruction techniques for the medial coll ateral ligament-deficient elbow." J Shoulder Elbow Surg 14(2): 207-15. Bisson, L. J., L. M. Manohar, et al. (2008) "Influence of suture material on the biomechanical behavior of suture-tendon speci mens: a controlled study in bovine rotator cuff." Am J Sports Med 36(5): 907-12. Brownhill, J. R., K. Furukawa, et al. (2006) "Surgeon accuracy in the selection of the flexion-extension axis of the elbow: an in vitro study." J Shoulder Elbow Surg 15(4): 451-6. Chhabra, A., J. S. Starman, et al. (2006). "Anatomic, radiographic, biomechanical, and kinematic evaluation of the an terior cruciate ligament and its two functional bundles." J Bone Joint Surg Am 88 Suppl 4: 2-10.

PAGE 102

91 Conway, J. E., F. W. Jobe, et al. (1992). "Medial instability of the elbow in throwing athletes. Treatment by repair or reconstruction of the ulnar co llateral ligament." J Bone Joint Surg Am 74(1): 67-83. Dumontier, C. (2006) Clinical Examinati on of the Elbow. Maitrise Orthopeddique Volume, DOI: Ellenbecker, T. S., A. J. Mattalino, et al. (1998). "Medial elbow joint laxity in professional baseball pitchers. A bilatera l comparison using stress radiography." Am J Sports Med 26(3): 420-4. Harrell, R. M., J. Tong, et al. (2003). "Compa rison of the mechanical properties of different tension band materials and suture techniques." J Orthop Trauma 17(2): 119-22. Hechtman, K. S., A. T. E. W. Tjin, et al (1998). "Biomechanics of a less invasive procedure for reconstruction of the ulnar coll ateral ligament of the elbow." Am J Sports Med 26(5): 620-4. Joseph A Buckwalter, T. E., Sheldon Simon ( 2000). Orthopaedic Basi c Science, Biology and Biomechanics of the Musculoskele tal system, The American Academy of Orthopaedic Surgeons. Jr., J. H. T. (2002). Biomechanical Evaluati on of the Intact, Inju red and Surgically Reconstructed Ulnar Collateral Ligament of the Elbow Joint mechanical engineering Philadelphia Drexel University. Master of Science in Mechanical Engineering: 147. Khuri, J., S. Wertheimer, et al. (2003). "Fixat ion of the offset V osteotomy: mechanical testing of 4 constructs." J Foot Ankle Surg 42(2): 63-7. Langer, P., P. Fadale, et al. (2006). "Evolution of the treatment options of ulnar collateral ligament injuries of the elbow." Br J Sports Med 40(6): 499-506. Lee, S. K., E. N. Kubiak, et al. (2005) "Thumb metacarpophalangeal ulnar collateral ligament injuries: a biomechanical simulation study of four static reconstructions." J Hand Surg [Am] 30(5): 1056-60. Leibman M., R. A. (2002). "Ulnar Collate ral Ligament." Medscape Orthopaedics & Sports Medicine 6(2): 1-3. Marieb, E. N. (2004). Joints. Human An atomy & Physiology, Pearson Benjamin Cummings: 258,271. McAdams, T. R., A. T. Lee, et al. (2007). "Two ulnar collateral ligament reconstruction methods: the docking technique versus bioa bsorbable interference screw fixation--a biomechanical evaluation with cyclic lo ading." J Shoulder Elbow Surg 16(2): 224-8.

PAGE 103

92 Miller, M. (1992). Review of Orthopaedics, W.B. Saunders Nassab, P. F. and M. S. Schickendantz (2006). "Evaluation and treatm ent of medial ulnar collateral ligament injuries in the throwing athlete." Sports Med Arthrosc 14(4): 221-31. Neal S. Elattrache, S. C. B., Tal Da vid (2001). "Medial Collateral Ligament Reconstruction Techniques in Should er & Elbow Surgery 2(1): 38-49. Paletta, G. A., Jr., S. J. Klepps, et al. ( 2006). "Biomechanical eval uation of 2 techniques for ulnar collateral ligament reconstructi on of the elbow." Am J Sports Med 34(10): 1599-603. Picard, F., A. M. DiGioia, et al. (2001 ). "Accuracy in tunnel placement for ACL reconstruction. Comparison of traditional ar throscopic and computer-assisted navigation techniques." Comput Aided Surg 6(5): 279-89. Pichora, J. E., K. Furukawa, et al. (2007). "I nitial repair strengths of two methods for acute medial collateral ligament injuries of the elbow." J Orthop Res 25(5): 612-6. Prud'homme, J., J. E. Budoff, et al. (2008). "B iomechanical analysis of medial collateral ligament reconstruction graf ts of the elbow." Am J Sports Med 36(4): 728-32. Rohrbough, J. T., D. W. Altchek, et al. (2002). "Medial collateral li gament reconstruction of the elbow using the docking techni que." Am J Sports Med 30(4): 541-8. Safran, M. R. and D. Baillargeon (2005). "Soft-tissue stabilizers of the elbow." J Shoulder Elbow Surg 14(1 Suppl S): 179S-185S. Shah, R. P.(2007) A Comparison of Cyclic Valgus Loading on Reconstructed Ulnar Collateral Ligament of the Yale University School of Medicine. Degree of Doctor of Medicine. 1-32 Tejwani, S. G., K. L. Markolf, et al (2005). "Reconstruction of the interosseous membrane of the forearm with a graft subs titute: a cadaveric study." J Hand Surg [Am] 30(2): 326-34.

PAGE 104

93 APPENDICES

PAGE 105

94 Appendix A: SAS Code data valgusstab; input recon cycle50 cycle100 cycle200 cycle600; cards ; 3 0.49 0.50 0.51 0.30 3 0.62 0.71 0.70 0.29 2 0.89 1.10 1.35 1.03 2 0.64 0.70 0.79 0.52 1 0.42 0.42 0.31 0.45 1 0.34 0.33 0.33 1.06 ; proc print data =valgusstab; run ; proc sort data =valgusstab; BY recon; run ; proc means data =valgusstab; BY recon; VAR cycle50 cycle100 cycle200 cycle600; run ; proc glm data =valgusstab; CLASS recon; MODEL cycle50 cycle100 cycle200 cycle600=recon; REPEATED cycle 4 ( 50 100 200 600 ); run ; means RECON/ TUKEY ; run ; data gapformation; input recon load1 load2 load3; cards ; 1 2.79 5.32 0.00 1 2.47 5.07 0.00 2 1.06 2.58 4.97 2 1.93 4.07 0.00 3 1.02 3.07 4.82 3 1.73 2.76 5.66 ; proc print data =gapformation; run ; proc sort data =gapformation; BY recon; run ; proc means data =gapformation; BY recon; VAR load1 load2 load3; run ; proc glm data =gapformation; CLASS recon; MODEL load1 load2 load3=recon; run ; means RECON/ TUKEY ;